Comparison of an electro-optical system and photo-conducting antenna employed as detectors of pulsed terahertz radiation by means of a new method for measuring spectral width
- ITMO University, St. Petersburg (Russian Federation)
Two detection systems, electro-optical system and photoconducting system, are tested by the method suggested previously for determining the boundaries of broadband terahertz radiation in time-domain spectroscopy. From a series of measurements the error in determining the operation ranges is calculated. The terahertz spectrometer with an electro-optical detector based on a ZnTe (110) crystal of thickness 2 mm has the operation spectral range of 0.059 – 1.092 THz. The detector utilizing an iPCA-21-05-1000-800-h photo-conducting antenna with the same source of signal demonstrates a wider operation band ranging from 0.017 to 1.6 THz. The method developed makes it possible to experimentally compare the parameters of the considered terahertz spectrometers obtained under the same quality of adjustment. (laser applications and other topics in quantum electronics)
- OSTI ID:
- 22373329
- Journal Information:
- Quantum Electronics (Woodbury, N.Y.), Vol. 44, Issue 12; Other Information: Country of input: International Atomic Energy Agency (IAEA); ISSN 1063-7818
- Country of Publication:
- United States
- Language:
- English
Similar Records
HIGH-SENSITIVITY AND BROADBAND TERAHERTZ FOCAL PLANE ARRAYS BASED ON PLASMONIC PHOTOCONDUCTIVE NANOCAVITIES
Evaluation of DAST and zinc telluride nonlinear crystals for efficient terahertz generation