Title: SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888840· OSTI ID:22369746
; ; ; ; ; ; ; ; ;  [1]
  1. UniversityAlabama Birmingham, Birmingham, AL (United Kingdom)

Purpose: HDR brachytherapy using interstitial needle template for cervical cancer is commonly delivered in 4-5 fractions. Routine verification of needle positions before each fraction is often based on radiographic imaging of implanted fiducial markers. The current study evaluated interfractional displacement of implanted fiducial markers using CT images. Methods: 9 sequential patients with cervical interstitial needle implants were evaluated. The superior and inferior borders of the target volumes were defined by fiducial markers in planning CT. The implant position was verified with kV orthogonal images before each fraction. A second CT was acquired prior 3rd fraction (one or 2 days post planning CT). Distances from inferior and superior fiducial markers to pubic symphysis plane (perpendicular to vaginal obtulator)were measured. Distance from needle tip of a reference needle (next to the inferior marker) to the pubic symphysis plane was also determined. The difference in fiducial marker distance or needle tip distance between planning CT and CT prior 3rd fraction were measured to assess markers migration and needle displacement. Results: The mean inferior marker displacement was 4.5 mm and ranged 0.9 to 11.3 mm. The mean superior marker displacement was 2.7 mm and ranged 0 to 10.4 mm. There was a good association between inferior and superior marker displacement (r=0.95). Mean averaged inferior and superior marker displacement was 3.3 mm and ranged from 0.1 to 10.9 mm, with a standard deviation of 3.2 mm. The mean needle displacement was 5.6 mm and ranged 0.2 to 15.6 mm. Needle displacements were reduced (p<0.05) after adjusting according to needle-to-fiducials distance. Conclusion: There were small fiducial marker displacements between HDR fractions. Our study suggests a target margin of 9.7 mm to cover interfractional marker displacements (in 95% cases) for pretreatment verification based on radiographic imaging.

OSTI ID:
22369746
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

Interfractional Position Variation of Pancreatic Tumors Quantified Using Intratumoral Fiducial Markers and Daily Cone Beam Computed Tomography
Journal Article · Sun Sep 01 00:00:00 EDT 2013 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22369746

Daily electronic portal imaging of implanted gold seed fiducials in patients undergoing radiotherapy after radical prostatectomy
Journal Article · Thu Feb 01 00:00:00 EST 2007 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22369746

Cervix Motion in 50 Cervical Cancer Patients Assessed by Daily Cone Beam Computed Tomographic Imaging of a New Type of Marker
Journal Article · Sun Nov 01 00:00:00 EDT 2015 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22369746