  6. University of North Carolina, Chapel Hill, NC (United States)

Purpose: Rotational step-and-shot IMRT (r-IMRT) could improve delivery efficiency with good dose conformity, especially if it can leverage the burst mode of the accelerator where radiation is turned on/off momentarily while the gantry rotates continuously. The challenge for the r-IMRT planning is to minimize the number of beams to achieve a fast and smooth rotational delivery. Methods: Treatment plans for r-IMRT were created using an in-house treatment planning system. To generate the plan using a very few beams, gantry angle was optimized by weighting the beam monitoring unit (MU), and beam shape optimization was a combination of column search with k-means clustering. A prostate case and a head and neck case were planned using r-IMRT. The dosimetry is compared to s-IMRT planned with Varian Eclipse treatment planning system. Results: With the same PTV dose coverage D95=100%, the r-IMRT plans shows comparable sparing as the s-IMRT plans in the prostate for the rectum D10cc and the bladder Dmean, and in the head and neck for the spinal cord Dmax, the brain stem Dmax, the left/right parotid Dmean, the larynx Dmean, and the mandible Dmean. Both plans meet the established institutional clinical dosimetric criteria. The r-IMRT plan uses 19 beam/405 MU for the prostate, and 68 beam/880 MU for the head and neck, while the s-IMRT uses 7 beam/724 MU and 9 beam/1812 MU, respectively. Compared to the corresponding s-IMRT, r-IMRT has a reduction of MUs of 44% for the prostate case and 41% for the head and neck case. Conclusions: We have demonstrated the feasibility of a rotational step and shoot IMRT treatment planning approach that significantly shortens the conventional IMRT treatment beam-on time without degrading the dose comformity.

OSTI ID:
22369648
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Feasibility of a unified approach to intensity-modulated radiation therapy and volume-modulated arc therapy optimization and delivery
Journal Article · Sat Feb 14 23:00:00 EST 2015 · Medical Physics · OSTI ID:22413438

Sci—Fri AM: Mountain — 05: Unified Optimization and Delivery of Intensity-modulated Radiation Therapy and Volume-modulated Arc Therapy
Journal Article · Fri Aug 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22407702

Many-isocenter optimization for robotic radiotherapy
Journal Article · Sun Feb 09 23:00:00 EST 2020 · Physics in Medicine and Biology (Online) · OSTI ID:1803104