skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: SU-E-T-463: Quantification of Rotational Variation in Mouse Setup for IGRT

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888796· OSTI ID:22369607
; ; ; ; ;  [1]
  1. UT MD Anderson Cancer Center, Houston, TX (United States)

Purpose: New small-animal irradiators include extremely precise IGRT capabilities. However, mouse immobilization and localization remains a challenge. In particular, unlike week-to-week translational displacements, rotational changes in positioning are not easily corrected for in subject setup. Using two methods of setup, we aim to quantify week-to-week rotational variation in mice for the purpose of IGRT planning in small animal studies. Methods: Ten mice were imaged weekly using breath-hold CBCT (X-RAD 225 Cx), with the mouse positioned in a half-pipe support, providing 40 scans. A second group of two mice were positioned in a 3D printed immobilization device, which was created using a CT from a similarly shaped mouse, providing 10 scans. For each mouse, the first image was taken to be the reference image. Subsequent CT images were then rigidly registered, based on bony anatomy. Rotations in the axial (roll), sagittal (pitch), and coronal (yaw) planes were recorded and used to quantify variation in angular setup. Results: For the mice imaged in the half pipe, average magnitude of roll was found to be 5.4±4.6° (range: −12.9:18.86°), of pitch 1.6±1.3° (range: −1.4:4.7°), and of yaw 1.9±1.5° (range −5.4:1.1°). For the mice imaged in the printed setup; average magnitude of roll was found to be 0.64±0.6° (range: −2.1:1.0°), of pitch 0.6±0.4° (range: 0.0:1.3°), and of yaw 0.2±0.1° (range: 0.0:0.4°). The printed setup provided reduction in roll, pitch, and yaw by 88, 62, and 90 percent, respectively. Conclusion: For the typical setup routine, roll in mouse position is the dominant source of rotational variation. However, when a printed device was used, drastic improvements in mouse immobilization were seen. This work provides a promising foundation for mouse immobilization, required for full scale small animal IGRT. Currently, we are making improvements to allo±w the use of a similar system for MR, PET, and bioluminescence.

OSTI ID:
22369607
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

SU-E-J-244: Validation of a 6D-Robotic-Couch and Image Guidance Radiation Therapy (IGRT) System for Localization in World's First Single-Room Proton Therapy System
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22369607

A Study on Target Positioning Error and Its Impact on Dose Variation in Image-Guided Stereotactic Body Radiotherapy for the Spine
Journal Article · Wed Apr 01 00:00:00 EDT 2009 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22369607

Pitch, roll, and yaw variations in patient positioning
Journal Article · Wed Nov 01 00:00:00 EST 2006 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22369607