SU-E-T-300: Spatial Variations of Multiple Off-Axial Targets for a Single Isocenter SRS Treatment Plan in ExacTrac 6D Robotic Couch System
- Genesis Medical Center, Davenport, IA (United States)
- Mount Sinai Hospital, NY, NY (United States)
Purpose: To evaluate the spatial variations of multiple off-axial targets for a single isocenter stereotactic radiosurgery (SRS) treatment plan in ExacTrac 6D robotic couch system (BrainLab AG). Methods: Five metallic ball bearing (BB) markers were placed sparsely in 3D off-axial locations (non-coplanar) inside a skull phantom as the representatives of multiple targets mimicking multiple brain metastases. The locations of the BB markers were carefully chosen to minimize overlapping of each other in a port imaging detector plane. The skull phantom was immobilized by a frameless mask and CT scanned with a BrainLab Head and Neck Localizer using a GE Optima MDCT scanner. The CT images were exported to iPlan software (BrainLab AG) and a multiple target PTV was drawn by combining all the contours of the BBs. The margin of the MLC opening was selected as 3 mm expansion outward. Two coplanar arc beams were placed to generate a single isocenter SRS plan to treat the PTV. The arc beams were delivered using Novalis Tx system with portal imaging acquisition mode per 10% temporal resolution. The locations of the BBs were visualized and analyzed with respect to the MLC aperture in the treatment plan similar to the Winston-Lutz test. Results: All the BBs were clearly identified inside the MLC openings. The positional errors for the BBs were overall less than 1 mm along the rotational path of the two arcs. Conclusion: This study verified that the spatial deviations of multiple off-axial targets for a single isocenter SRS treatment plan is within sub-millimeter range in ExacTrac 6D robotic couch system. Accompanied with the Winston-Lutz test, this test will quality-assure the spatial accuracies of the isocenter as well as the positions of multiple off-axial targets for the SRS treatment using a single isocenter multiple target treatment plan.
- OSTI ID:
- 22355860
- Journal Information:
- Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-E-J-123: Targeting Accuracy of Image-Guided Radiosurgery for Intracranial Lesions
SU-F-J-177: A Novel Image Analysis Technique (center Pixel Method) to Quantify End-To-End Tests