SU-E-J-150: Impact of Intrafractional Prostate Motion On the Accuracy and Efficiency of Prostate SBRT Delivery: A Retrospective Analysis of Prostate Tracking Log Files

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888203· OSTI ID:22334042
; ; ;  [1]; ; ; ;  [2];  [3]; ; ;  [1]
  1. Massachusetts General Hospital and Harvard Medical School, Boston, MA (United States)
  2. Boston Medical Center and Boston University School of Medicine, Boston, MA (United States)
  3. Massachusetts Institute of Technology, Cambridge, MA (United States)
Purpose: To measure intrafractional prostate motion by time-based stereotactic x-ray imaging and investigate the impact on the accuracy and efficiency of prostate SBRT delivery. Methods: Prostate tracking log files with 1,892 x-ray image registrations from 18 SBRT fractions for 6 patients were retrospectively analyzed. Patient setup and beam delivery sessions were reviewed to identify extended periods of large prostate motion that caused delays in setup or interruptions in beam delivery. The 6D prostate motions were compared to the clinically used PTV margin of 3–5 mm (3 mm posterior, 5 mm all other directions), a hypothetical PTV margin of 2–3 mm (2 mm posterior, 3 mm all other directions), and the rotation correction limits (roll ±2°, pitch ±5° and yaw ±3°) of CyberKnife to quantify beam delivery accuracy. Results: Significant incidents of treatment start delay and beam delivery interruption were observed, mostly related to large pitch rotations of ≥±5°. Optimal setup time of 5–15 minutes was recorded in 61% of the fractions, and optimal beam delivery time of 30–40 minutes in 67% of the fractions. At a default imaging interval of 15 seconds, the percentage of prostate motion beyond PTV margin of 3–5 mm varied among patients, with a mean at 12.8% (range 0.0%–31.1%); and the percentage beyond PTV margin of 2–3 mm was at a mean of 36.0% (range 3.3%–83.1%). These timely detected offsets were all corrected real-time by the robotic manipulator or by operator intervention at the time of treatment interruptions. Conclusion: The durations of patient setup and beam delivery were directly affected by the occurrence of large prostate motion. Frequent imaging of down to 15 second interval is necessary for certain patients. Techniques for reducing prostate motion, such as using endorectal balloon, can be considered to assure consistently higher accuracy and efficiency of prostate SBRT delivery.
OSTI ID:
22334042
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

TU-F-CAMPUS-J-01: Inference of Prostate PTV Margins in VMAT Delivery From Intra-Fraction Prostate Motion During SBRT Delivery
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22570035

Continuous Monitoring and Intrafraction Target Position Correction During Treatment Improves Target Coverage for Patients Undergoing SBRT Prostate Therapy
Journal Article · Sat Feb 28 23:00:00 EST 2015 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22458629

Real-Time Study of Prostate Intrafraction Motion During External Beam Radiotherapy With Daily Endorectal Balloon
Journal Article · Wed Nov 30 23:00:00 EST 2011 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22054474