Title: SU-E-I-18: CT Scanner QA Using Normalized CTDI Ratio

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4887966· OSTI ID:22325137
; ;  [1]
  1. San Diego State University, San Diego, CA (United States)

Purpose: To create a ratio of weighted computed tomography dose index (CTDIw) data normalized to in-air measurements (CTDIair) as a function of beam quality to create a look-up table for frequent, rapid quality assurance (QA) checks of CTDI. Methods: The CTDIw values were measured according to TG-63 protocol using a pencil ionization chamber (Unfors Xi CT detector) and head and body Polymethyl methacrylate (PMMA) phantoms (16 and 32 cm diameter, respectively). Single scan dose profiles were measured at each clinically available energy (80,100,120,140 kVp) on three different CT scanners (two Siemens SOMATOM Definition Flash and one GE Optima), using a tube current of 400 mA, a one second rotation time, and the widest available beam width (32 × 0.6 mm and 16 × 1.25 mm, respectively). These values were normalized to CTDIair measurements using the same conditions as CTDIw. The ratios (expressed in cGy/R) were assessed for each scanner as a function of each energy's half value layer (HVL) paired with the phantom's appropriate bow tie filter measured in mmAl. Results: Normalized CTDI values vary linearly with HVL for both the head and body phantoms. The ratios for the two Siemens machines are very similar at each energy. Compared to the GE scanner, these values vary between 10–20% for each kVp setting. Differences in CTDIair contribute most to the deviation of the ratios across machines. Ratios are independent of both mAs and collimation. Conclusion: Look-up tables constructed of normalized CTDI values as a function of HVL can be used to derive CTDIw data from only three in-air measurements (one for CTDIair and two with added filtration for HVL) to allow for simple, frequent QA checks without CT phantom setup. Future investigations will involve comparing results with Monte Carlo simulations for validation.

OSTI ID:
22325137
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English