skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Non-contact direct measurement of the magnetocaloric effect in thin samples

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4890394· OSTI ID:22308689
;  [1]
  1. Department of Physics and Earth Sciences, University of Parma, Parco Area delle Scienze 7/A, 43124 Parma (Italy)

An experimental setup, based on a non-contact temperature sensor, is proposed to directly measure the magnetocaloric effect of samples few micrometers thick. The measurement of the adiabatic temperature change of foils and ribbons is fundamental to design innovative devices based on magnetocaloric thin materials or micro-structuring bulk samples. The reliability of the proposed setup is demonstrated by comparing the measurements performed on a bulk gadolinium sample with the results obtained by an experimental setup based on a Cernox bare chip thermoresistance and by in-field differential scanning calorimetry. We show that this technique can measure the adiabatic temperature variation on gadolinium sheets as thin as 27 μm. Heat transfer simulations are added to describe the capability of the presented technique.

OSTI ID:
22308689
Journal Information:
Review of Scientific Instruments, Vol. 85, Issue 7; Other Information: (c) 2014 AIP Publishing LLC; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English