Signal evaluations using singular value decomposition for Thomson scattering diagnostics
Journal Article
·
· Review of Scientific Instruments
- Japan Atomic Energy Agency, 801-1 Mukoyama, Naka 311-0193 (Japan)
- National Institute for Fusion Science, 322-6 Oroshi-cho, Toki 509-5292 (Japan)
This paper provides a novel method for evaluating signal intensities in incoherent Thomson scattering diagnostics. A double-pass Thomson scattering system, where a laser passes through the plasma twice, generates two scattering pulses from the plasma. Evaluations of the signal intensities in the spectrometer are sometimes difficult due to noise and stray light. We apply the singular value decomposition method to Thomson scattering data with strong noise components. Results show that the average accuracy of the measured electron temperature (T{sub e}) is superior to that of temperature obtained using a low-pass filter (<20 MHz) or without any filters.
- OSTI ID:
- 22303471
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 85; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Stray light analysis for the Thomson scattering diagnostic of the ETE Tokamak
Thomson scattering diagnostic system design for the Compact Toroidal Hybrid experiment
A simple, high performance Thomson scattering diagnostic for high temperature plasma research
Journal Article
·
Tue Dec 14 23:00:00 EST 2010
· Review of Scientific Instruments
·
OSTI ID:22062215
Thomson scattering diagnostic system design for the Compact Toroidal Hybrid experiment
Journal Article
·
Fri Nov 14 23:00:00 EST 2014
· Review of Scientific Instruments
·
OSTI ID:22303698
A simple, high performance Thomson scattering diagnostic for high temperature plasma research
Technical Report
·
Mon Jan 31 23:00:00 EST 1994
·
OSTI ID:10142389