A comparison of HDR near source dosimetry using a treatment planning system, Monte Carlo simulation, and radiochromic film
Journal Article
·
· Medical Dosimetry
- Department of Radiation Oncology, University of Cincinnati College of Medicine, Cincinnati, OH (United States)
This study aimed to investigate the high-dose rate Iridium-192 brachytherapy, including near source dosimetry, of a catheter-based applicator from 0.5 mm to 1 cm along the transverse axis. Radiochromic film and Monte Carlo (MC) simulation were used to generate absolute dose for the catheter-based applicator. Results from radiochromic film and MC simulation were compared directly to the treatment planning system (TPS) based on the American Association of Physicists in Medicine Updated Task Group 43 (TG-43U1) dose calculation formalism. The difference between dose measured using radiochromic film along the transverse plane at 0.5 mm from the surface and the predicted dose by the TPS was 24%±13%. The dose difference between the MC simulation along the transverse plane at 0.5 mm from the surface and the predicted dose by the TPS was 22.1%±3%. For distances from 1.5 mm to 1 cm from the surface, radiochromic film and MC simulation agreed with TPS within an uncertainty of 3%. The TPS under-predicts the dose at the surface of the applicator, i.e., 0.5 mm from the catheter surface, as compared to the measured and MC simulation predicted dose. MC simulation results demonstrated that 15% of this error is due to neglecting the beta particles and discrete electrons emanating from the sources and not considered by the TPS, and 7% of the difference was due to the photon alone, potentially due to the differences in MC dose modeling, photon spectrum, scoring techniques, and effect of the presence of the catheter and the air gap. Beyond 1 mm from the surface, the TPS dose algorithm agrees with the experimental and MC data within 3%.
- OSTI ID:
- 22262807
- Journal Information:
- Medical Dosimetry, Journal Name: Medical Dosimetry Journal Issue: 2 Vol. 38; ISSN 0958-3947
- Country of Publication:
- United States
- Language:
- English
Similar Records
Proposed linear energy transfer areal detector for protons using radiochromic film
Design and evaluation of a HDR skin applicator with flattening filter
Radiochromic film dosimetry of HDR {sup 192}Ir source radiation fields
Journal Article
·
Wed Apr 15 00:00:00 EDT 2015
· Review of Scientific Instruments
·
OSTI ID:22392464
Design and evaluation of a HDR skin applicator with flattening filter
Journal Article
·
Thu Feb 14 23:00:00 EST 2008
· Medical Physics
·
OSTI ID:21036153
Radiochromic film dosimetry of HDR {sup 192}Ir source radiation fields
Journal Article
·
Mon Nov 14 23:00:00 EST 2011
· Medical Physics
·
OSTI ID:22098669