Evaluation of 4-dimensional Computed Tomography to 4-dimensional Cone-Beam Computed Tomography Deformable Image Registration for Lung Cancer Adaptive Radiation Therapy
Journal Article
·
· International Journal of Radiation Oncology, Biology and Physics
- Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia (United States)
- Department of Electrical and Computer Engineering, University of Iowa, Iowa City, Iowa (United States)
- Department of Radiation Oncology, University of Iowa, Iowa City, Iowa (United States)
- Department of Radiation Oncology, State University of New York Upstate Medical University, Syracuse, New York (United States)
- Sydney Medical School, University of Sydney, Sydney, New South Wales (Australia)
Purpose: To evaluate 2 deformable image registration (DIR) algorithms for the purpose of contour mapping to support image-guided adaptive radiation therapy with 4-dimensional cone-beam CT (4DCBCT). Methods and Materials: One planning 4D fan-beam CT (4DFBCT) and 7 weekly 4DCBCT scans were acquired for 10 locally advanced non-small cell lung cancer patients. The gross tumor volume was delineated by a physician in all 4D images. End-of-inspiration phase planning 4DFBCT was registered to the corresponding phase in weekly 4DCBCT images for day-to-day registrations. For phase-to-phase registration, the end-of-inspiration phase from each 4D image was registered to the end-of-expiration phase. Two DIR algorithms—small deformation inverse consistent linear elastic (SICLE) and Insight Toolkit diffeomorphic demons (DEMONS)—were evaluated. Physician-delineated contours were compared with the warped contours by using the Dice similarity coefficient (DSC), average symmetric distance, and false-positive and false-negative indices. The DIR results are compared with rigid registration of tumor. Results: For day-to-day registrations, the mean DSC was 0.75 ± 0.09 with SICLE, 0.70 ± 0.12 with DEMONS, 0.66 ± 0.12 with rigid-tumor registration, and 0.60 ± 0.14 with rigid-bone registration. Results were comparable to intraobserver variability calculated from phase-to-phase registrations as well as measured interobserver variation for 1 patient. SICLE and DEMONS, when compared with rigid-bone (4.1 mm) and rigid-tumor (3.6 mm) registration, respectively reduced the average symmetric distance to 2.6 and 3.3 mm. On average, SICLE and DEMONS increased the DSC to 0.80 and 0.79, respectively, compared with rigid-tumor (0.78) registrations for 4DCBCT phase-to-phase registrations. Conclusions: Deformable image registration achieved comparable accuracy to reported interobserver delineation variability and higher accuracy than rigid-tumor registration. Deformable image registration performance varied with the algorithm and the patient.
- OSTI ID:
- 22224486
- Journal Information:
- International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 2 Vol. 86; ISSN IOBPD3; ISSN 0360-3016
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-F-J-84: Comparison of Quantitative Deformable Image Registration Evaluation Tools: Application to Prostate IGART
SU-E-J-151: Dosimetric Evaluation of DIR Mapped Contours for Image Guided Adaptive Radiotherapy with 4D Cone-Beam CT
SU-F-J-88: Comparison of Two Deformable Image Registration Algorithms for CT-To-CT Contour Propagation
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22632209
SU-E-J-151: Dosimetric Evaluation of DIR Mapped Contours for Image Guided Adaptive Radiotherapy with 4D Cone-Beam CT
Journal Article
·
Sun Jun 01 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22334043
SU-F-J-88: Comparison of Two Deformable Image Registration Algorithms for CT-To-CT Contour Propagation
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22632213