Title: Comparison of the Effects of High-Energy Photon Beam Irradiation (10 and 18 MV) on 2 Types of Implantable Cardioverter-Defibrillators

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1]; ; ; ;  [1]; ;  [2];  [3]; ;  [1]
  1. Department of Radiation Oncology, University of Tsukuba, Tsukuba, Ibaraki (Japan)
  2. Department of Radiation Oncology, Shizuoka Cancer Center Hospital, Nagaizumi, Shizuoka (Japan)
  3. Department of Cardiovascular Medicine, University of Tsukuba, Tsukuba, Ibaraki (Japan)

Purpose: Radiation therapy for cancer may be required for patients with implantable cardiac devices. However, the influence of secondary neutrons or scattered irradiation from high-energy photons (≥10 MV) on implantable cardioverter-defibrillators (ICDs) is unclear. This study was performed to examine this issue in 2 ICD models. Methods and Materials: ICDs were positioned around a water phantom under conditions simulating clinical radiation therapy. The ICDs were not irradiated directly. A control ICD was positioned 140 cm from the irradiation isocenter. Fractional irradiation was performed with 18-MV and 10-MV photon beams to give cumulative in-field doses of 600 Gy and 1600 Gy, respectively. Errors were checked after each fraction. Soft errors were defined as severe (change to safety back-up mode), moderate (memory interference, no changes in device parameters), and minor (slight memory change, undetectable by computer). Results: Hard errors were not observed. For the older ICD model, the incidences of severe, moderate, and minor soft errors at 18 MV were 0.75, 0.5, and 0.83/50 Gy at the isocenter. The corresponding data for 10 MV were 0.094, 0.063, and 0 /50 Gy. For the newer ICD model at 18 MV, these data were 0.083, 2.3, and 5.8 /50 Gy. Moderate and minor errors occurred at 18 MV in control ICDs placed 140 cm from the isocenter. The error incidences were 0, 1, and 0 /600 Gy at the isocenter for the newer model, and 0, 1, and 6 /600Gy for the older model. At 10 MV, no errors occurred in control ICDs. Conclusions: ICD errors occurred more frequently at 18 MV irradiation, which suggests that the errors were mainly caused by secondary neutrons. Soft errors of ICDs were observed with high energy photon beams, but most were not critical in the newer model. These errors may occur even when the device is far from the irradiation field.

OSTI ID:
22224390
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 85, Issue 3; Other Information: Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English