Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Efficacy and Safety of a Novel Vascular Closure Device (Glubran 2 Seal) After Diagnostic and Interventional Angiography in Patients with Peripheral Arterial Occlusive Disease

Journal Article · · Cardiovascular and Interventional Radiology
 [1]; ; ;  [2];  [1];  [2]; ;  [1];  [2];  [1];  [2]
  1. University of Pisa, Department of Vascular Surgery (Italy)
  2. University of Pisa, Department of Diagnostic and Interventional Radiology (Italy)
To prospectively evaluate safety and efficacy of a novel vascular closure device (Glubran 2 Seal) after peripheral angiography in patients with peripheral arterial occlusive disease (PAOD). From December 2010 to June 2011, all consecutive patients with PAOD undergoing peripheral angiography were prospectively enrolled onto the study after percutaneous antegrade or retrograde puncture of the common femoral artery. After angiography, the Glubran 2 Seal device was used to achieve hemostasis. The following data were registered: technical success and manual compression duration, patients' discomfort (scale 0-5), operators' technical difficulty (scale 0-5), and vascular complications. The site of hemostasis was evaluated by clinical inspection and color-coded Duplex ultrasound performed 1 day and 1 month after the procedure. One hundred seventy-eight patients were enrolled (112 male, mean age 70.8 years) with a total of 206 puncture sites, including 104 (50.5 %) antegrade accesses. The device was successful in 198(96.1 %) of 206 procedures, with 8 cases of manual compression lasting longer than 5 min (maximum 20 min). No major vascular complications were observed, resulting in 100 % procedural success. Minor complications occurred in seven procedures (3.4 %), including two cases of pseudoaneurysms, successfully treated by ultrasound-guided glue injection. The mean {+-} standard deviation score for patients' discomfort was 0.9 {+-} 0.7, whereas the mean score for operators' difficulty was 1.2 {+-} 0.9. In patients with PAOD, the Glubran 2 Seal represents a simple, painless, and efficient vascular closure device, able to achieve hemostasis both in antegrade and retrograde accesses.
OSTI ID:
22156377
Journal Information:
Cardiovascular and Interventional Radiology, Journal Name: Cardiovascular and Interventional Radiology Journal Issue: 2 Vol. 36; ISSN 0174-1551; ISSN CAIRDG
Country of Publication:
United States
Language:
English

Similar Records

Suitability of Exoseal Vascular Closure Device for Antegrade Femoral Artery Puncture Site Closure
Journal Article · Sat Jun 15 00:00:00 EDT 2013 · Cardiovascular and Interventional Radiology · OSTI ID:22113646

The Use of ExoSeal Vascular Closure Device for Direct Antegrade Superficial Femoral Artery Puncture Site Hemostasis
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Cardiovascular and Interventional Radiology · OSTI ID:22469923

Prospective Nonrandomized Trial of Manual Compression and Angio-Seal and Starclose Arterial Closure Devices in Common Femoral Punctures
Journal Article · Sun Apr 15 00:00:00 EDT 2007 · Cardiovascular and Interventional Radiology · OSTI ID:21091014