Intrafraction Verification of Gated RapidArc by Using Beam-Level Kilovoltage X-Ray Images

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
; ; ; ; ; ; ;  [1]
  1. Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California (United States)

Purpose: To verify the geometric accuracy of gated RapidArc treatment using kV images acquired during dose delivery. Methods and Materials: Twenty patients were treated using the gated RapidArc technique with a Varian TrueBeam STx linear accelerator. One to 7 metallic fiducial markers were implanted inside or near the tumor target before treatment simulation. For patient setup and treatment verification purposes, the internal target volume (ITV) was created, corresponding to each implanted marker. The gating signal was generated from the Real-time Position Management (RPM) system. At the beginning of each fraction, individualized respiratory gating amplitude thresholds were set based on fluoroscopic image guidance. During the treatment, we acquired kV images immediately before MV beam-on at every breathing cycle, using the on-board imaging system. After the treatment, all implanted markers were detected, and their 3-dimensional (3D) positions in the patient were estimated using software developed in-house. The distance from the marker to the corresponding ITV was calculated for each patient by averaging over all markers and all fractions. Results: The average 3D distance between the markers and their ITVs was 0.8 {+-} 0.5 mm (range, 0-1.7 mm) and was 2.1 {+-} 1.2 mm at the 95th percentile (range, 0-3.8 mm). On average, a left-right margin of 0.6 mm, an anterior-posterior margin of 0.8 mm, and a superior-inferior margin of 1.5 mm is required to account for 95% of the intrafraction uncertainty in RPM-based RapidArc gating. Conclusion: To our knowledge, this is the first clinical report of intrafraction verification of respiration-gated RapidArc treatment in stereotactic ablative radiation therapy. For some patients, the markers deviated significantly from the ITV by more than 2 mm at the beginning of the MV beam-on. This emphasizes the need for gating techniques with beam-on/-off controlled directly by the actual position of the tumor target instead of external surrogates such as RPM.

OSTI ID:
22149433
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 5 Vol. 83; ISSN IOBPD3; ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

Evaluation of the geometric accuracy of surrogate-based gated VMAT using intrafraction kilovoltage x-ray images
Journal Article · Tue May 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22098853

Kilovoltage Intrafraction Monitoring for Prostate Intensity Modulated Arc Therapy: First Clinical Results
Journal Article · Fri Nov 30 23:00:00 EST 2012 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:22149694

Comparison of Kilovoltage Cone-Beam Computed Tomography With Megavoltage Projection Pairs for Paraspinal Radiosurgery Patient Alignment and Position Verification
Journal Article · Fri Aug 01 00:00:00 EDT 2008 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21124416