Title: Determining thyroid {sup 131}I effective half-life for the treatment planning of Graves' disease

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4788660· OSTI ID:22130536
; ; ;  [1];  [2]
  1. Cancer Institute of Sao Paulo State (ICESP), Clinical Hospital, School of Medicine, University of Sao Paulo, Sao Paulo 01246-000 (Brazil)
  2. Unit of Medical Physics, Azienda Ospedaliero-Universitaria Pisana, Pisa 56126 (Italy)

Purpose: Thyroid {sup 131}I effective half-life (T{sub eff}) is an essential parameter in patient therapy when accurate radiation dose is desirable for producing an intended therapeutic outcome. Multiple {sup 131}I uptake measurements and resources from patients themselves and from nuclear medicine facilities are requisites for determining T{sub eff}, these being limiting factors when implementing the treatment planning of Graves' disease (GD) in radionuclide therapy. With the aim of optimizing this process, this study presents a practical, propitious, and accurate method of determining T{sub eff} for dosimetric purposes. Methods: A total of 50 patients with GD were included in this prospective study. Thyroidal {sup 131}I uptake was measured at 2-h, 6-h, 24-h, 48-h, 96-h, and 220-h postradioiodine administration. T{sub eff} was calculated by considering sets of two measured points (24-48-h, 24-96-h, and 24-220-h), sets of three (24-48-96-h, 24-48-220-h, and 24-96-220-h), and sets of four (24-48-96-220-h). Results: When considering all the measured points, the representative T{sub eff} for all the patients was 6.95 ({+-}0.81) days, whereas when using such sets of points as (24-220-h), (24-96-220-h), and (24-48-220-h), this was 6.85 ({+-}0.81), 6.90 ({+-}0.81), and 6.95 ({+-}0.81) days, respectively. According to the mean deviations 2.2 ({+-}2.4)%, 2.1 ({+-}2.0)%, and 0.04 ({+-}0.09)% found in T{sub eff}, calculated based on all the measured points in time, and with methods using the (24-220-h), (24-48-220-h), and (24-96-220-h) sets, respectively, no meaningful statistical difference was noted among the three methods (p > 0.500, t test). Conclusions: T{sub eff} obtained from only two thyroid {sup 131}I uptakes measured at 24-h and 220-h, besides proving to be sufficient, accurate enough, and easily applicable, attributes additional major cost-benefits for patients, and facilitates the application of the method for dosimetric purposes in the treatment planning of Graves' disease.

OSTI ID:
22130536
Journal Information:
Medical Physics, Vol. 40, Issue 2; Other Information: (c) 2013 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English