Time resolution of a 1-inch cylindrical CeBr{sub 3} crystal at {sup 60}Co energies
Journal Article
·
· AIP Conference Proceedings
- Grupo de Fisica Nuclear, Universidad Complutense, Madrid (Spain)
- Grupo de Fisica Nuclear, Universidad Complutense, Madrid, Spain and Medical and Scientific Time Imaging Consulting, Nykoeping (Sweden)
We have measured time resolutions of a cylindrical CeBr{sub 3} scintillator of 1-inch in height and 1-inch in diameter coupled to two different fast photomultiplier tubes, Hamamatsu R9779 and Photonis XP20D0, as a function of applied high voltages and different settings of a Constant Fraction Discriminator ORTEC 935. The time resolution was measured using a time-delayed coincidence set-up involving a fast reference detector. The best result of 119(2) ps at {sup 60}Co energies was obtained for the CeBr{sub 3} crystal coupled to the Hamamatsu PMT. This result is comparable to the resolution of 107 ps reported for a LaBr{sub 3}(Ce) crystal of the same size. For the coupling of the CeBr{sub 3} scintillator to the Photonis PMT we got the time resolution of 146(2) ps.
- OSTI ID:
- 22118602
- Journal Information:
- AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 1541; ISSN APCPCS; ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Characterization of the Hamamatsu 8-inch R14688-100 PMT
Time Resolution of Fast Photomultipliers for Time of Flight PET
High spatial resolution gamma imaging detector based on a 5 inch diameter R3292 Hamamatsu PSPMT
Journal Article
·
Sun Feb 25 19:00:00 EST 2024
· Journal of Instrumentation
·
OSTI ID:2406445
Time Resolution of Fast Photomultipliers for Time of Flight PET
Journal Article
·
Mon Jan 04 23:00:00 EST 2010
· AIP Conference Proceedings
·
OSTI ID:21366902
High spatial resolution gamma imaging detector based on a 5 inch diameter R3292 Hamamatsu PSPMT
Journal Article
·
Mon Jun 01 00:00:00 EDT 1998
· IEEE Transactions on Nuclear Science
·
OSTI ID:624172
Related Subjects
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
CERIUM BROMIDES
COBALT 60
COMPARATIVE EVALUATIONS
CRYSTALS
CYLINDRICAL CONFIGURATION
DISCRIMINATORS
ELECTRIC POTENTIAL
GAMMA DETECTION
LANTHANUM BROMIDES
MEV RANGE 01-10
PHOTOMULTIPLIERS
READOUT SYSTEMS
SOLID SCINTILLATION DETECTORS
TIME DELAY
TIME RESOLUTION
CERIUM BROMIDES
COBALT 60
COMPARATIVE EVALUATIONS
CRYSTALS
CYLINDRICAL CONFIGURATION
DISCRIMINATORS
ELECTRIC POTENTIAL
GAMMA DETECTION
LANTHANUM BROMIDES
MEV RANGE 01-10
PHOTOMULTIPLIERS
READOUT SYSTEMS
SOLID SCINTILLATION DETECTORS
TIME DELAY
TIME RESOLUTION