Title: Dosimetric characterization of round HDR {sup 192}Ir AccuBoost applicators for breast brachytherapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3232001· OSTI ID:22102140
; ; ;  [1]
  1. Department of Radiation Oncology, Tufts University School of Medicine, Boston, Massachusetts 02111 (United States)

Purpose: The AccuBoost brachytherapy system applies HDR {sup 192}Ir beams peripherally to the breast using collimating applicators. The purpose of this study was to benchmark Monte Carlo simulations of the HDR {sup 192}Ir source, to dosimetrically characterize the round applicators using established Monte Carlo simulation and radiation measurement techniques and to gather data for clinical use. Methods: Dosimetric measurements were performed in a polystyrene phantom, while simulations estimated dose in air, liquid water, polystyrene and ICRU 44 breast tissue. Dose distribution characterization of the 4-8 cm diameter collimators was performed using radiochromic EBT film and air ionization chambers. Results: The central axis dose falloff was steeper for the 4 cm diameter applicator in comparison to the 8 cm diameter applicator, with surface to 3 cm depth-dose ratios of 3.65 and 2.44, respectively. These ratios did not considerably change when varying the phantom composition from breast tissue to polystyrene, phantom thickness from 4 to 8 cm, or phantom radius from 8 to 15 cm. Dose distributions on the central axis were fitted to sixth-order polynomials for clinical use in a hand calculation spreadsheet (i.e., nomogram). Dose uniformity within the useful applicator apertures decreased as depth-dose increased. Conclusions: Monte Carlo benchmarking simulations of the HDR {sup 192}Ir source using the MCNP5 radiation transport code indicated agreement within 1% of the published results over the radial/angular region of interest. Changes in phantom size and radius did not cause noteworthy changes in the central axis depth-dose. Polynomial fit depth-dose curves provide a simple and accurate basis for a nomogram.

OSTI ID:
22102140
Journal Information:
Medical Physics, Vol. 36, Issue 11; Other Information: (c) 2009 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English