Title: Ionization chamber gradient effects in nonstandard beam configurations

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3213518· OSTI ID:22102115
; ; ;  [1]
  1. Departement de Physique, Universite de Montreal, Pavillon Roger-Gaudry (D-428), 2900 Boulevard Edouard-Montpetit, Montreal, Quebec H3T 1J4, Canada and Departement de Radio-oncologie, Centre Hospitalier de l'Universite de Montreal (CHUM), 1560 Sherbrooke est, Montreal, Quebec H2L 4M1 (Canada)

Purpose: For the purpose of nonstandard beam reference dosimetry, the current concept of reporting absorbed dose at a point in water located at a representative position in the chamber volume is investigated in detail. As new nonstandard beam reference dosimetry protocols are under development, an evaluation of the role played by the definition of point of measurement could lead to conceptual improvements prior to establishing measurement procedures. Methods: The present study uses the current definition of reporting absorbed dose to calculate ionization chamber perturbation factors for two cylindrical chamber models (Exradin A12 and A14) using the Monte Carlo method. The EGSnrc based user-code EGS lowbar chamber is used to calculate chamber dose responses of 14 IMRT beams chosen to cause considerable dose gradients over the chamber volume as previously used by Bouchard and Seuntjens [''Ionization chamber-based reference dosimetry of intensity modulated radiation beams,'' Med. Phys. 31(9), 2454-5465 (2004)]. Results: The study shows conclusively the relative importance of each physical effect involved in the nonstandard beam correction factors of 14 IMRT beams. Of all correction factors involved in the dosimetry of the beams studied, the gradient perturbation correction factor has the highest magnitude, on average, 11% higher compared to reference conditions for the Exradin A12 chamber and about 5% higher for the Extradin A14 chamber. Other perturbation correction factors (i.e., P{sub wall}, P{sub stem}, and P{sub cel}) are, on average, less than 0.8% different from reference conditions for the chambers and beams studied. The current approach of reporting measured absorbed dose at a point in water coinciding with the location of the centroid of the chamber is the main factor responsible for large correction factors in nonstandard beam deliveries (e.g., intensity modulated radiation therapy) reported in literature. Conclusions: To reduce or eliminate the magnitude of perturbation correction factors in nonstandard beam reference dosimetry, two possible ways to report absorbed dose are suggested: (1) Reporting average dose to the sensitive volume of the chamber filled with water, combined with removing the reference field implicit gradient effect when measuring output factors, and (2) reporting average dose to the chamber itself during output factor verifications. The first option could be adopted if clinical beam correction factors are negligible. The second option could simplify quality assurance procedures when correction factors are not negligible and have to be calculated using Monte Carlo simulations.

OSTI ID:
22102115
Journal Information:
Medical Physics, Vol. 36, Issue 10; Other Information: (c) 2009 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English