Systematic uncertainties in the Monte Carlo calculation of ion chamber replacement correction factors
- Ottawa Carleton Institute of Physics, Carleton University, Campus Ottawa, Ottawa, Ontario KIS 5B6 (Canada)
In a previous study [Med. Phys. 35, 1747-1755 (2008)], the authors proposed two direct methods of calculating the replacement correction factors (P{sub repl} or p{sub cav}p{sub dis}) for ion chambers by Monte Carlo calculation. By ''direct'' we meant the stopping-power ratio evaluation is not necessary. The two methods were named as the high-density air (HDA) and low-density water (LDW) methods. Although the accuracy of these methods was briefly discussed, it turns out that the assumption made regarding the dose in an HDA slab as a function of slab thickness is not correct. This issue is reinvestigated in the current study, and the accuracy of the LDW method applied to ion chambers in a {sup 60}Co photon beam is also studied. It is found that the two direct methods are in fact not completely independent of the stopping-power ratio of the two materials involved. There is an implicit dependence of the calculated P{sub repl} values upon the stopping-power ratio evaluation through the choice of an appropriate energy cutoff {Delta}, which characterizes a cavity size in the Spencer-Attix cavity theory. Since the {Delta} value is not accurately defined in the theory, this dependence on the stopping-power ratio results in a systematic uncertainty on the calculated P{sub repl} values. For phantom materials of similar effective atomic number to air, such as water and graphite, this systematic uncertainty is at most 0.2% for most commonly used chambers for either electron or photon beams. This uncertainty level is good enough for current ion chamber dosimetry, and the merits of the two direct methods of calculating P{sub repl} values are maintained, i.e., there is no need to do a separate stopping-power ratio calculation. For high-Z materials, the inherent uncertainty would make it practically impossible to calculate reliable P{sub repl} values using the two direct methods.
- OSTI ID:
- 22100511
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 5 Vol. 36; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Accuracy of Spencer-Attix cavity theory and calculations of fluence correction factors for the air kerma formalism
Calculation of the replacement correction factors for ion chambers in megavoltage beams by Monte Carlo simulation
Modified energy-deposition model, for the computation of the stopping-power ratio for small cavity sizes
Journal Article
·
Tue Sep 15 00:00:00 EDT 2009
· Medical Physics
·
OSTI ID:22102088
Calculation of the replacement correction factors for ion chambers in megavoltage beams by Monte Carlo simulation
Journal Article
·
Thu May 15 00:00:00 EDT 2008
· Medical Physics
·
OSTI ID:21120681
Modified energy-deposition model, for the computation of the stopping-power ratio for small cavity sizes
Journal Article
·
Sat Feb 28 23:00:00 EST 1981
· Phys. Rev., A; (United States)
·
OSTI ID:6609154