Attenuation-based estimation of patient size for the purpose of size specific dose estimation in CT. Part II. Implementation on abdomen and thorax phantoms using cross sectional CT images and scanned projection radiograph images

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4757586· OSTI ID:22099080
; ; ; ; ;  [1]
  1. Department of Radiology, Mayo Clinic, Rochester, Minnesota 55905 (United States)

Purpose: To estimate attenuation using cross sectional CT images and scanned projection radiograph (SPR) images in a series of thorax and abdomen phantoms. Methods: Attenuation was quantified in terms of a water cylinder with cross sectional area of A{sub w} from both the CT and SPR images of abdomen and thorax phantoms, where A{sub w} is the area of a water cylinder that would absorb the same dose as the specified phantom. SPR and axial CT images were acquired using a dual-source CT scanner operated at 120 kV in single-source mode. To use the SPR image for estimating A{sub w}, the pixel values of a SPR image were calibrated to physical water attenuation using a series of water phantoms. A{sub w} and the corresponding diameter D{sub w} were calculated using the derived attenuation-based methods (from either CT or SPR image). A{sub w} was also calculated using only geometrical dimensions of the phantoms (anterior-posterior and lateral dimensions or cross sectional area). Results: For abdomen phantoms, the geometry-based and attenuation-based methods gave similar results for D{sub w}. Using only geometric parameters, an overestimation of D{sub w} ranging from 4.3% to 21.5% was found for thorax phantoms. Results for D{sub w} using the CT image and SPR based methods agreed with each other within 4% on average in both thorax and abdomen phantoms. Conclusions: Either the cross sectional CT or SPR images can be used to estimate patient attenuation in CT. Both are more accurate than use of only geometrical information for the task of quantifying patient attenuation. The SPR based method requires calibration of SPR pixel values to physical water attenuation and this calibration would be best performed by the scanner manufacturer.

OSTI ID:
22099080
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 11 Vol. 39; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Application- and patient size-dependent optimization of x-ray spectra for CT
Journal Article · Sun Mar 15 00:00:00 EDT 2009 · Medical Physics · OSTI ID:22098471

Development of a method to estimate organ doses for pediatric CT examinations
Journal Article · Sun May 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22620872

Traceable micro-CT scaling accuracy phantom for applications requiring exact measurement of distances or volumes
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Medical Physics · OSTI ID:22099031