Title: Dosimetric measurements of Onyx embolization material for stereotactic radiosurgery

Abstract

Purpose: Arteriovenous malformations are often treated with a combination of embolization and stereotactic radiosurgery. Concern has been expressed in the past regarding the dosimetric properties of materials used in embolization and the effects that the introduction of these materials into the brain may have on the quality of the radiosurgery plan. To quantify these effects, the authors have taken large volumes of Onyx 34 and Onyx 18 (ethylene-vinyl alcohol copolymer doped with tantalum) and measured the attenuation and interface effects of these embolization materials. Methods: The manufacturer provided large cured volumes ({approx}28 cc) of both Onyx materials. These samples were 8.5 cm in diameter with a nominal thickness of 5 mm. The samples were placed on a block tray above a stack of solid water with an Attix chamber at a depth of 5 cm within the stack. The Attix chamber was used to measure the attenuation. These measurements were made for both 6 and 16 MV beams. Placing the sample directly on the solid water stack and varying the thickness of solid water between the sample and the Attix chamber measured the interface effects. The computed tomography (CT) numbers for bulk material were measured in a phantom using amore » wide bore CT scanner. Results: The transmission through the Onyx materials relative to solid water was approximately 98% and 97% for 16 and 6 MV beams, respectively. The interface effect shows an enhancement of approximately 2% and 1% downstream for 16 and 6 MV beams. CT numbers of approximately 2600-3000 were measured for both materials, which corresponded to an apparent relative electron density (RED) {rho}{sub e}{sup w} to water of approximately 2.7-2.9 if calculated from the commissioning data of the CT scanner. Conclusions: We performed direct measurements of attenuation and interface effects of Onyx 34 and Onyx 18 embolization materials with large samples. The introduction of embolization materials affects the dose distribution of a MV therapeutic beam, but should be of negligible consequence for effective thicknesses of less than 8 mm. The measured interface effects are also small, particularly at 6 MV. Large areas of high-density artifacts and low-density artifacts can cause errors in dose calculations and need to be identified and resolved during planning.« less

Authors:
; ; ; ;  [1];  [2];  [2];  [2]
  1. Radiation Physics Division, Department of Radiation Oncology, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109 (United States)
  2. (United States)
Publication Date:
OSTI Identifier:
22099077
Resource Type:
Journal Article
Journal Name:
Medical Physics
Additional Journal Information:
Journal Volume: 39; Journal Issue: 11; Other Information: (c) 2012 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); Journal ID: ISSN 0094-2405
Country of Publication:
United States
Language:
English
Subject:
62 RADIOLOGY AND NUCLEAR MEDICINE; 61 RADIATION PROTECTION AND DOSIMETRY; 60 APPLIED LIFE SCIENCES; BRAIN; COMPUTERIZED TOMOGRAPHY; COPOLYMERS; DOPED MATERIALS; ELECTRON DENSITY; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; SURGERY; WATER

Citation Formats

Roberts, Donald A., Balter, James M., Chaudhary, Neeraj, Gemmete, Joseph J., Pandey, Aditya S., Division of Neurointerventional Radiology, Departments of Radiology and Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109, Division of Neurointerventional Radiology, Departments of Radiology, Neurosurgery, and Otolaryngology, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109, and Department of Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109. Dosimetric measurements of Onyx embolization material for stereotactic radiosurgery. United States: N. p., 2012. Web. doi:10.1118/1.4757918.
Roberts, Donald A., Balter, James M., Chaudhary, Neeraj, Gemmete, Joseph J., Pandey, Aditya S., Division of Neurointerventional Radiology, Departments of Radiology and Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109, Division of Neurointerventional Radiology, Departments of Radiology, Neurosurgery, and Otolaryngology, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109, & Department of Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109. Dosimetric measurements of Onyx embolization material for stereotactic radiosurgery. United States. doi:10.1118/1.4757918.
Roberts, Donald A., Balter, James M., Chaudhary, Neeraj, Gemmete, Joseph J., Pandey, Aditya S., Division of Neurointerventional Radiology, Departments of Radiology and Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109, Division of Neurointerventional Radiology, Departments of Radiology, Neurosurgery, and Otolaryngology, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109, and Department of Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109. Thu . "Dosimetric measurements of Onyx embolization material for stereotactic radiosurgery". United States. doi:10.1118/1.4757918.
@article{osti_22099077,
title = {Dosimetric measurements of Onyx embolization material for stereotactic radiosurgery},
author = {Roberts, Donald A. and Balter, James M. and Chaudhary, Neeraj and Gemmete, Joseph J. and Pandey, Aditya S. and Division of Neurointerventional Radiology, Departments of Radiology and Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109 and Division of Neurointerventional Radiology, Departments of Radiology, Neurosurgery, and Otolaryngology, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109 and Department of Neurosurgery, University of Michigan Hospital and Health Systems, 1500 East Medical Center Drive, Ann Arbor, Michigan 48109},
abstractNote = {Purpose: Arteriovenous malformations are often treated with a combination of embolization and stereotactic radiosurgery. Concern has been expressed in the past regarding the dosimetric properties of materials used in embolization and the effects that the introduction of these materials into the brain may have on the quality of the radiosurgery plan. To quantify these effects, the authors have taken large volumes of Onyx 34 and Onyx 18 (ethylene-vinyl alcohol copolymer doped with tantalum) and measured the attenuation and interface effects of these embolization materials. Methods: The manufacturer provided large cured volumes ({approx}28 cc) of both Onyx materials. These samples were 8.5 cm in diameter with a nominal thickness of 5 mm. The samples were placed on a block tray above a stack of solid water with an Attix chamber at a depth of 5 cm within the stack. The Attix chamber was used to measure the attenuation. These measurements were made for both 6 and 16 MV beams. Placing the sample directly on the solid water stack and varying the thickness of solid water between the sample and the Attix chamber measured the interface effects. The computed tomography (CT) numbers for bulk material were measured in a phantom using a wide bore CT scanner. Results: The transmission through the Onyx materials relative to solid water was approximately 98% and 97% for 16 and 6 MV beams, respectively. The interface effect shows an enhancement of approximately 2% and 1% downstream for 16 and 6 MV beams. CT numbers of approximately 2600-3000 were measured for both materials, which corresponded to an apparent relative electron density (RED) {rho}{sub e}{sup w} to water of approximately 2.7-2.9 if calculated from the commissioning data of the CT scanner. Conclusions: We performed direct measurements of attenuation and interface effects of Onyx 34 and Onyx 18 embolization materials with large samples. The introduction of embolization materials affects the dose distribution of a MV therapeutic beam, but should be of negligible consequence for effective thicknesses of less than 8 mm. The measured interface effects are also small, particularly at 6 MV. Large areas of high-density artifacts and low-density artifacts can cause errors in dose calculations and need to be identified and resolved during planning.},
doi = {10.1118/1.4757918},
journal = {Medical Physics},
issn = {0094-2405},
number = 11,
volume = 39,
place = {United States},
year = {2012},
month = {11}
}