Commissioning a CT-compatible LDR tandem and ovoid applicator using Monte Carlo calculation and 3D dosimetry
- Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina 27710 (United States)
Purpose: To determine the geometric and dose attenuation characteristics of a new commercially available CT-compatible LDR tandem and ovoid (T and O) applicator using Monte Carlo calculation and 3D dosimetry. Methods: For geometric characterization, we quantified physical dimensions and investigated a systematic difference found to exist between nominal ovoid angle and the angle at which the afterloading buckets fall within the ovoid. For dosimetric characterization, we determined source attenuation through asymmetric gold shielding in the buckets using Monte Carlo simulations and 3D dosimetry. Monte Carlo code MCNP5 was used to simulate 1.5 Multiplication-Sign 10{sup 9} photon histories from a {sup 137}Cs source placed in the bucket to achieve statistical uncertainty of 1% at a 6 cm distance. For 3D dosimetry, the distribution about an unshielded source was first measured to evaluate the system for {sup 137}Cs, after which the distribution was measured about sources placed in each bucket. Cylindrical PRESAGE{sup Registered-Sign} dosimeters (9.5 cm diameter, 9.2 cm height) with a central channel bored for source placement were supplied by Heuris Inc. The dosimeters were scanned with the Duke Large field of view Optical CT-Scanner before and after delivering a nominal dose at 1 cm of 5-8 Gy. During irradiation the dosimeter was placed in a water phantom to provide backscatter. Optical CT scan time lasted 15 min during which 720 projections were acquired at 0.5 Degree-Sign increments, and a 3D distribution was reconstructed with a (0.05 cm){sup 3} isotropic voxel size. The distributions about the buckets were used to calculate a 3D distribution of transmission rate through the bucket, which was applied to a clinical CT-based T and O implant plan. Results: The systematic difference in bucket angle relative to the nominal ovoid angle (105 Degree-Sign ) was 3.1 Degree-Sign -4.7 Degree-Sign . A systematic difference in bucket angle of 1 Degree-Sign , 5 Degree-Sign , and 10 Degree-Sign caused a 1%{+-} 0.1%, 1.7%{+-} 0.4%, and 2.6%{+-} 0.7% increase in rectal dose, respectively, with smaller effect to dose to Point A, bladder, sigmoid, and bowel. For 3D dosimetry, 90.6% of voxels had a 3D {gamma}-index (criteria = 0.1 cm, 3% local signal) below 1.0 when comparing measured and expected dose about the unshielded source. Dose transmission through the gold shielding at a radial distance of 1 cm was 85.9%{+-} 0.2%, 83.4%{+-} 0.7%, and 82.5%{+-} 2.2% for Monte Carlo, and measurement for left and right buckets, respectively. Dose transmission was lowest at oblique angles from the bucket with a minimum of 56.7%{+-} 0.8%, 65.6%{+-} 1.7%, and 57.5%{+-} 1.6%, respectively. For a clinical T and O plan, attenuation from the buckets leads to a decrease in average Point A dose of {approx}3.2% and decrease in D{sub 2cc} to bladder, rectum, bowel, and sigmoid of 5%, 18%, 6%, and 12%, respectively. Conclusions: Differences between dummy and afterloading bucket position in the ovoids is minor compared to effects from asymmetric ovoid shielding, for which rectal dose is most affected. 3D dosimetry can fulfill a novel role in verifying Monte Carlo calculations of complex dose distributions as are common about brachytherapy sources and applicators.
- OSTI ID:
- 22098932
- Journal Information:
- Medical Physics, Vol. 39, Issue 7; Other Information: (c) 2012 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Development of prototype shielded cervical intracavitary brachytherapy applicators compatible with CT and MR imaging
Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE{sup Registered-Sign} radiochromic plastic
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
AFTERLOADING
BLADDER
BRACHYTHERAPY
CAT SCANNING
CESIUM 137
COMPUTERIZED SIMULATION
DOSEMETERS
DOSIMETRY
IRRADIATION
MONTE CARLO METHOD
PHANTOMS
PHOTONS
RADIATION DOSE DISTRIBUTIONS
RADIATION DOSES
RADIATION SOURCE IMPLANTS
RECTUM
SHIELDING