Title: Characteristics of optically stimulated luminescence dosimeters in the spread-out Bragg peak region of clinical proton beams

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3693055· OSTI ID:22098804
; ;  [1]
  1. Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, Texas 77030 (United States)

Purpose: Optically stimulated luminescent detectors (OSLDs) have a number of advantages in radiation dosimetry making them excellent dosimeters for quality assurance and patient dose verification. Although the dosimeters have been investigated in several modalities, relatively little work has been done in examining the dosimeters for use in clinical proton beams. This study examined a number of characteristics of the response of the dosimeters in the spread-out Bragg peak (SOBP) region of clinical proton beams. Methods: Optically stimulated luminescence (OSL) dosimeters from Landauer, Inc., specifically the nanoDot dosimeter, were investigated. These dosimeters were placed in a special phantom with a recess to fit the dosimeters without an air gap. Beams with nominal energies of 160, 200, and 250 MeV were used in the passively-scattered proton beam at the MD Anderson Cancer Center Proton Therapy Center. Dosimetric properties including linearity, field size dependence, energy dependence, residual signal as a function of cumulative dose, and postirradiation fading were investigated by taking measurements at the center of SOBPs. Results: The dosimeters showed 1% supralinearity at 200 cGy and 5% supralinearity at 1000 cGy. No noticeable field size dependence of the detector was found for field sizes from 2 x 2 cm{sup 2} to 18 x 18 cm{sup 2}. Residual signal as a function of cumulative dose showed a small increase for measurements up to 1000 cGy. Readout signal depletion of the dosimeters after consecutive readings showed a slightly larger depletion in protons for doses up to 500 cGy but not by a clinically significant amount. Within the center of various SOBP widths and proton energies the variation in response was less than 2%. An average beam quality factor of 1.089 with experimental standard deviation of 0.007 was determined and applied to the data such that the results were within 1.2% of ion chamber data. Conclusions: The nanoDot OSL dosimeter characteristics were studied in the SOBP region of clinical proton beams. To achieve accurate dosimetric readings, corrections to the dosimeter response were applied. Corrections tended to be minimal or broadly consistent. The nanoDot OSLD was found to be an acceptable dosimeter for measurement in the SOBP region for a range of clinical proton beams.

OSTI ID:
22098804
Journal Information:
Medical Physics, Vol. 39, Issue 4; Other Information: (c) 2012 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English