Title: Reference dosimetry for helical tomotherapy: Practical implementation and a multicenter validation

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3651496· OSTI ID:22098667
; ; ; ;  [1]
  1. ZNA-Middelheim, Department Radiotherapy Lindendreef 1, 2020 Antwerpen (Belgium)

Purpose: The aim of this study was to implement a protocol for reference dosimetry in tomotherapy and to validate the beam output measurements with an independent dosimetry system. Methods: Beam output was measured at the reference depth of 10 cm in water for the following three cases: (1) a 5 x 10 cm{sup 2} static machine specific reference field (MSR), (2) a rotational 5 x 10 cm{sup 2} field without modulation and no tabletop in the beam, (3) a plan class specific reference (PCSR) field defined as a rotational homogeneous dose delivery to a cylindrical shaped target volume: plan with modulation and table-top movement. The formalism for reference dosimetry of small and nonstandard fields [Med.Phys.35: 5179-5186, 2008] and QA recommendations [Med.Phys.37: 4817-4853, 2010] were adopted in the dose measurement protocol. All ionization chamber measurements were verified independently using alanine/EPR dosimetry. As a pilot study, the beam output was measured on tomotherapy Hi-art systems at three other centers and directly compared to the centers specifications and to alanine dosimetry. Results: For the four centers, the mean static output at a depth of 10 cm in water and SAD = 85 cm, measured with an A1SL chamber following the TG-148 report was 6.238 Gy/min {+-} 0.058 (1 SD); the rotational output was 6.255 Gy/min {+-} 0.069 (1 SD). The dose stated by the center was found in good agreement with the measurements of the visiting team: D{sub center}/D{sub visit} = 1.000 {+-} 0.003 (1 SD). The A1SL chamber measurements were all in good agreement with Alanine/EPR dosimetry. Going from the static reference field to the rotational/non modulated field the dose rate remains constant within 0.2% except for one center where a deviation of 1.3% was detected. Conclusions: Following the TG-148 report, beam output measurements in water at the reference depth using a local protocol, as developed at different centers, was verified. The measurements were found in good agreement with alanine/EPR dosimetry. The presented methodology may provide a good concept for reference dosimetry.

OSTI ID:
22098667
Journal Information:
Medical Physics, Vol. 38, Issue 11; Other Information: (c) 2011 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English