Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA
- Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 53705 (United States)
Purpose: The purpose of this work is to explore the usefulness of the gamma passing rate metric for per-patient, pretreatment dose QA and to validate a novel patient-dose/DVH-based method and its accuracy and correlation. Specifically, correlations between: (1) gamma passing rates for three 3D dosimeter detector geometries vs clinically relevant patient DVH-based metrics; (2) Gamma passing rates of whole patient dose grids vs DVH-based metrics, (3) gamma passing rates filtered by region of interest (ROI) vs DVH-based metrics, and (4) the capability of a novel software algorithm that estimates corrected patient Dose-DVH based on conventional phan-tom QA data are analyzed. Methods: Ninety six unique ''imperfect'' step-and-shoot IMRT plans were generated by applying four different types of errors on 24 clinical Head/Neck patients. The 3D patient doses as well as the dose to a cylindrical QA phantom were then recalculated using an error-free beam model to serve as a simulated measurement for comparison. Resulting deviations to the planned vs simulated measured DVH-based metrics were generated, as were gamma passing rates for a variety of difference/distance criteria covering: dose-in-phantom comparisons and dose-in-patient comparisons, with the in-patient results calculated both over the whole grid and per-ROI volume. Finally, patient dose and DVH were predicted using the conventional per-beam planar data as input into a commercial ''planned dose perturbation'' (PDP) algorithm, and the results of these predicted DVH-based metrics were compared to the known values. Results: A range of weak to moderate correlations were found between clinically relevant patient DVH metrics (CTV-D95, parotid D{sub mean}, spinal cord D1cc, and larynx D{sub mean}) and both 3D detector and 3D patient gamma passing rate (3%/3 mm, 2%/2 mm) for dose-in-phantom along with dose-in-patient for both whole patient volume and filtered per-ROI. There was considerable scatter in the gamma passing rate vs DVH-based metric curves. However, for the same input data, the PDP estimates were in agreement with actual patient DVH results. Conclusions: Gamma passing rate, even if calculated based on patient dose grids, has generally weak correlation to critical patient DVH errors. However, the PDP algorithm was shown to accurately predict the DVH impact using conventional planar QA results. Using patient-DVH-based metrics IMRT QA allows per-patient dose QA to be based on metrics that are both sensitive and specific. Further studies are now required to analyze new processes and action levels associated with DVH-based metrics to ensure effectiveness and practicality in the clinical setting.
- OSTI ID:
- 22098640
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 10 Vol. 38; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors
A Quality Assurance Method that Utilizes 3D Dosimetry and Facilitates Clinical Interpretation
MO-FG-CAMPUS-TeP1-01: An Efficient Method of 3D Patient Dose Reconstruction Based On EPID Measurements for Pre-Treatment Patient Specific QA
Journal Article
·
Mon Feb 14 23:00:00 EST 2011
· Medical Physics
·
OSTI ID:22096891
A Quality Assurance Method that Utilizes 3D Dosimetry and Facilitates Clinical Interpretation
Journal Article
·
Mon Oct 01 00:00:00 EDT 2012
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:22149543
MO-FG-CAMPUS-TeP1-01: An Efficient Method of 3D Patient Dose Reconstruction Based On EPID Measurements for Pre-Treatment Patient Specific QA
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22653893
Related Subjects
61 RADIATION PROTECTION AND DOSIMETRY
62 RADIOLOGY AND NUCLEAR MEDICINE
ACCURACY
ALGORITHMS
CORRELATIONS
DISTURBANCES
DOSIMETRY
ERRORS
GAMMA RADIATION
LARYNX
METRICS
MODULATION
PATIENTS
PERTURBATION THEORY
PHANTOMS
QUALITY ASSURANCE
RADIATION DOSES
RADIATION MONITORING
RADIATION PROTECTION
RADIOTHERAPY
SIMULATION
62 RADIOLOGY AND NUCLEAR MEDICINE
ACCURACY
ALGORITHMS
CORRELATIONS
DISTURBANCES
DOSIMETRY
ERRORS
GAMMA RADIATION
LARYNX
METRICS
MODULATION
PATIENTS
PERTURBATION THEORY
PHANTOMS
QUALITY ASSURANCE
RADIATION DOSES
RADIATION MONITORING
RADIATION PROTECTION
RADIOTHERAPY
SIMULATION