A planning and delivery study of a rotational IMRT technique with burst delivery
- Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, Wisconsin 53226 (United States)
Purpose: A novel rotational IMRT (rIMRT) technique using burst delivery (continuous gantry rotation with beam off during MLC repositioning) is investigated. The authors evaluate the plan quality and delivery efficiency and accuracy of this dynamic technique with a conventional flat 6 MV photon beam. Methods: Burst-delivery rIMRT was implemented in a planning system and delivered with a 160-MLC linac. Ten rIMRT plans were generated for five anonymized patient cases encompassing head and neck, brain, prostate, and prone breast. All plans were analyzed retrospectively and not used for treatment. Among the varied plan parameters were the number of optimization points, number of arcs, gantry speed, and gantry angle range (alpha) over which the beam is turned on at each optimization point. Combined rotational/step-and-shoot rIMRT plans were also created by superimposing multiple-segment static fields at several optimization points. The rIMRT trial plans were compared with each other and with plans generated using helical tomotherapy and VMAT. Burst-mode rotational IMRT plans were delivered and verified using a diode array, ionization chambers, thermoluminescent dosimeters, and film. Results: Burst-mode rIMRT can achieve plan quality comparable to helical tomotherapy, while the former may lead to slightly better OAR sparing for certain cases and the latter generally achieves slightly lower hot spots. Few instances were found in which increasing the number of optimization points above 36, or superimposing step-and-shoot IMRT segments, led to statistically significant improvements in OAR sparing. Using an additional rIMRT partial arc yielded substantial OAR dose improvements for the brain case. Measured doses from the rIMRT plan delivery were within 4% of the plan calculation in low dose gradient regions. Delivery time range was 228-375 s for single-arc rIMRT 200-cGy prescription with a 300 MU/min dose rate, comparable to tomotherapy and VMAT. Conclusions: Rotational IMRT with burst delivery, whether combined with static fields or not, yields clinically acceptable and deliverable treatment plans.
- OSTI ID:
- 22098615
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 9 Vol. 38; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Volumetric modulated arc therapy: IMRT in a single gantry arc
Assessing the Role of Volumetric Modulated Arc Therapy (VMAT) Relative to IMRT and Helical Tomotherapy in the Management of Localized, Locally Advanced, and Post-Operative Prostate Cancer
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
BRAIN
COMPUTERIZED TOMOGRAPHY
CT-GUIDED RADIOTHERAPY
DELIVERY
EFFICIENCY
FILMS
GYNECOLOGY
HOT SPOTS
IONIZATION CHAMBERS
LINEAR ACCELERATORS
MAMMARY GLANDS
PATIENTS
PHOTON BEAMS
PLANNING
PROSTATE
QUALITY ASSURANCE
RADIATION DOSES
RADIATION PROTECTION
ROTATION