Title: Measured and Monte Carlo calculated k{sub Q} factors: Accuracy and comparison

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3600697· OSTI ID:22098591
; ;  [1]
  1. Ottawa Medical Physics Institute (OMPI), Ottawa Carleton Institute for Physics, Carleton University Campus, 1125 Colonel By Drive, Ottawa, Ontario K1S 5B6 (Canada)

Purpose: The journal Medical Physics recently published two papers that determine beam quality conversion factors, k{sub Q}, for large sets of ion chambers. In the first paper [McEwen Med. Phys. 37, 2179-2193 (2010)], k{sub Q} was determined experimentally, while the second paper [Muir and Rogers Med. Phys. 37, 5939-5950 (2010)] provides k{sub Q} factors calculated using Monte Carlo simulations. This work investigates a variety of additional consistency checks to verify the accuracy of the k{sub Q} factors determined in each publication and a comparison of the two data sets. Uncertainty introduced in calculated k{sub Q} factors by possible variation of W/e with beam energy is investigated further. Methods: The validity of the experimental set of k{sub Q} factors relies on the accuracy of the NE2571 reference chamber measurements to which k{sub Q} factors for all other ion chambers are correlated. The stability of NE2571 absorbed dose to water calibration coefficients is determined and comparison to other experimental k{sub Q} factors is analyzed. Reliability of Monte Carlo calculated k{sub Q} factors is assessed through comparison to other publications that provide Monte Carlo calculations of k{sub Q} as well as an analysis of the sleeve effect, the effect of cavity length and self-consistencies between graphite-walled Farmer-chambers. Comparison between the two data sets is given in terms of the percent difference between the k{sub Q} factors presented in both publications. Results: Monitoring of the absorbed dose calibration coefficients for the NE2571 chambers over a period of more than 15 yrs exhibit consistency at a level better than 0.1%. Agreement of the NE2571 k{sub Q} factors with a quadratic fit to all other experimental data from standards labs for the same chamber is observed within 0.3%. Monte Carlo calculated k{sub Q} factors are in good agreement with most other Monte Carlo calculated k{sub Q} factors. Expected results are observed for the sleeve effect and the effect of cavity length on k{sub Q}. The mean percent differences between experimental and Monte Carlo calculated k{sub Q} factors are -0.08, -0.07, and -0.23% for the Elekta 6, 10, and 25 MV nominal beam energies, respectively. An upper limit on the variation of W/e in photon beams from cobalt-60 to 25 MV is determined as 0.4% with 95% confidence. The combined uncertainty on Monte Carlo calculated k{sub Q} factors is reassessed and amounts to between 0.40 and 0.49% depending on the wall material of the chamber. Conclusions: Excellent agreement (mean percent difference of only 0.13% for the entire data set) between experimental and calculated k{sub Q} factors is observed. For some chambers, k{sub Q} is measured for only one chamber of each type--the level of agreement observed in this study would suggest that for those chambers the measured k{sub Q} values are generally representative of the chamber type.

OSTI ID:
22098591
Journal Information:
Medical Physics, Vol. 38, Issue 8; Other Information: (c) 2011 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English