Radiation protection measurements around a 12 MeV mobile dedicated IORT accelerator
- Laboratory of Medical Physics, Istituto Regina Elena, via Elio Chianesi 53, 00144 Rome (Italy)
Purpose: The aim of this study is to investigate radioprotection issues that must be addressed when dedicated accelerators for intraoperative radiotherapy (IORT) are used in operating rooms. Recently, a new version of a mobile IORT accelerator (LIAC Sordina SpA, Italy) with 12 MeV electron beam has been implemented. This energy is necessary in some specific pathology treatments to allow a better coverage of thick lesions. At an electron energy of 10 MeV, leakage and scattered x-ray radiation (stray radiation) coming from the accelerator device and patient must be considered. If the energy is greater than 10 MeV, the x-ray component will increase; however, the most meaningful change should be the addition of neutron background. Therefore, radiation exposure of personnel during the IORT procedure needs to be carefully evaluated. Methods: In this study, stray x-ray radiation was measured and characterized in a series of spherical projections by means of an ion chamber survey meter. To simulate the patient during all measurements, a polymethylmethacrylate (PMMA) slab phantom with volume 30x30x15 cm{sup 3} and density 1.19 g/cm{sup 3} was used. The PMMA phantom was placed along the central axis of the beam in order to absorb the electron beams and the tenth value layer (TVL) and half value layer (HVL) of scattered radiation (at 0 deg., 90 deg., and 180 deg. scattering angles) were also measured at 1 m of distance from the phantom center. Neutron measurements were performed using passive bubble dosimeters and a neutron probe, specially designed to evaluate ambient dose equivalent H{sup *}(10). Results: The x-ray equivalent dose measured at 1 m along the beam axis at 12 MeV was 260 {mu}Sv/Gy. The value measured at 1 m at 90 deg. scattering angle was 25 {mu}Sv/Gy. The HVL and TVL values were 1.1 and 3.5 cm of lead at 0 deg., and 0.4 and 1 cm at 90 deg., respectively. The highest equivalent dose of fast neutrons was found to be at the surface of the phantom on the central beam axis (2.9{+-}0.6 {mu}Sv/Gy), while a lower value was observed below the phantom (1.6{+-}0.3 {mu}Sv/Gy). The neutron dose equivalent at 90 deg. scattering angle and on the floor plane on the beam axis below the beam stopper was negligible. Conclusions: Our data confirm that neutron exposure levels around the new dedicated IORT accelerator are very low. Mobile shielding panels can be used to reduce x-ray levels to below regulatory levels without necessarily providing permanent shielding in the operating room.
- OSTI ID:
- 22098508
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 3 Vol. 37; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Monte Carlo simulation estimates of neutron doses to critical organs of a patient undergoing 18 MV x-ray LINAC-based radiotherapy
Measurement of LINAC 90 deg. head leakage radiation TVL values
Measurement of the neutron leakage from a dedicated intraoperative radiation therapy electron linear accelerator and a conventional linear accelerator for 9, 12, 15(16), and 18(20) MeV electron energies
Journal Article
·
Wed Dec 14 23:00:00 EST 2005
· Medical Physics
·
OSTI ID:20726857
Measurement of LINAC 90 deg. head leakage radiation TVL values
Journal Article
·
Fri Sep 15 00:00:00 EDT 2006
· Medical Physics
·
OSTI ID:20853472
Measurement of the neutron leakage from a dedicated intraoperative radiation therapy electron linear accelerator and a conventional linear accelerator for 9, 12, 15(16), and 18(20) MeV electron energies
Journal Article
·
Thu May 15 00:00:00 EDT 2008
· Medical Physics
·
OSTI ID:21120679
Related Subjects
61 RADIATION PROTECTION AND DOSIMETRY
62 RADIOLOGY AND NUCLEAR MEDICINE
DOSE EQUIVALENTS
DOSEMETERS
DOSIMETRY
ELECTRON BEAMS
FAST NEUTRONS
IONIZATION CHAMBERS
LINEAR ACCELERATORS
MEDICAL PERSONNEL
NEUTRON BEAMS
PATHOLOGY
PHANTOMS
PMMA
RADIATION PROTECTION
RADIOTHERAPY
SHIELDING
STRAY RADIATION
X RADIATION
62 RADIOLOGY AND NUCLEAR MEDICINE
DOSE EQUIVALENTS
DOSEMETERS
DOSIMETRY
ELECTRON BEAMS
FAST NEUTRONS
IONIZATION CHAMBERS
LINEAR ACCELERATORS
MEDICAL PERSONNEL
NEUTRON BEAMS
PATHOLOGY
PHANTOMS
PMMA
RADIATION PROTECTION
RADIOTHERAPY
SHIELDING
STRAY RADIATION
X RADIATION