Field size effect of radiation quality in carbon therapy using passive method
- Advanced Applied Science Department, Research Laboratory, IHI Corporation, 1 Shin-Nakahara-Cho, Isogo-Ku, Yokohama 235-8501 (Japan)
The authors have investigated the dependency of radiation quality and absorbed dose on radiation field size in therapeutic carbon beams. The field size of the broad beam, formed using the passive technique, was controlled from 20 to 100 mm per side with a multileaf collimator. The absorbed dose and radiation quality on the beam center were evaluated at several depths in a water phantom using microdosimetric technique in experiments and Monte Carlo simulations. With an increase in the field size, the radiation quality was reduced, although the absorbed dose grew at the center of the field. This indicates that the dose and radiation quality at the center of the broad beam are influenced by particles from the off-center region via large-angle scattering and that such particles have relatively low radiation quality and mainly consist of fragment particles. Because such a tendency appeared to be more remarkable in the deeper region of the water phantom, it is likely that fragment particles that are born in a water phantom have a marked role in determining the field size effect.
- OSTI ID:
- 22098463
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 3 Vol. 36; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Dose contributions from large-angle scattered particles in therapeutic carbon beams
Dosimetry protocol for the forthcoming clinical trials in synchrotron stereotactic radiation therapy (SSRT)
Microdosimetric study for secondary neutrons in phantom produced by a 290 MeV/nucleon carbon beam
Journal Article
·
Sun Jan 14 23:00:00 EST 2007
· Medical Physics
·
OSTI ID:20853909
Dosimetry protocol for the forthcoming clinical trials in synchrotron stereotactic radiation therapy (SSRT)
Journal Article
·
Tue Mar 15 00:00:00 EDT 2011
· Medical Physics
·
OSTI ID:22096952
Microdosimetric study for secondary neutrons in phantom produced by a 290 MeV/nucleon carbon beam
Journal Article
·
Sat Sep 15 00:00:00 EDT 2007
· Medical Physics
·
OSTI ID:21032693