Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Measurement of compressed breast thickness by optical stereoscopic photogrammetry

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3065066· OSTI ID:22098436
; ;  [1]
  1. Room S657, Imaging Research, Sunnybrook Health Sciences Centre, 2075 Bayview Avenue, Toronto, Ontario M4N 3M5 (Canada)

The determination of volumetric breast density (VBD) from mammograms requires accurate knowledge of the thickness of the compressed breast. In attempting to accurately determine VBD from images obtained on conventional mammography systems, the authors found that the thickness reported by a number of mammography systems in the field varied by as much as 15 mm when compressing the same breast or phantom. In order to evaluate the behavior of mammographic compression systems and to be able to predict the thickness at different locations in the breast on patients, they have developed a method for measuring the local thickness of the breast at all points of contact with the compression paddle using optical stereoscopic photogrammetry. On both flat (solid) and compressible phantoms, the measurements were accurate to better than 1 mm with a precision of 0.2 mm. In a pilot study, this method was used to measure thickness on 108 volunteers who were undergoing mammography examination. This measurement tool will allow us to characterize paddle surface deformations, deflections and calibration offsets for mammographic units.

OSTI ID:
22098436
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 2 Vol. 36; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English