Dosimetric validation of Acuros XB with Monte Carlo methods for photon dose calculations
- Department of Medical Physics, British Columbia Cancer Agency-Vancouver Island Center, Victoria, British Columbia V8R 6V5 (Canada)
Purpose: The dosimetric accuracy of the recently released Acuros XB advanced dose calculation algorithm (Varian Medical Systems, Palo Alto, CA) is investigated for single radiation fields incident on homogeneous and heterogeneous geometries, and a comparison is made to the analytical anisotropic algorithm (AAA). Methods: Ion chamber measurements for the 6 and 18 MV beams within a range of field sizes (from 4.0x4.0 to 30.0x30.0 cm{sup 2}) are used to validate Acuros XB dose calculations within a unit density phantom. The dosimetric accuracy of Acuros XB in the presence of lung, low-density lung, air, and bone is determined using BEAMnrc/DOSXYZnrc calculations as a benchmark. Calculations using the AAA are included for reference to a current superposition/convolution standard. Results: Basic open field tests in a homogeneous phantom reveal an Acuros XB agreement with measurement to within {+-}1.9% in the inner field region for all field sizes and energies. Calculations on a heterogeneous interface phantom were found to agree with Monte Carlo calculations to within {+-}2.0%({sigma}{sub MC}=0.8%) in lung ({rho}=0.24 g cm{sup -3}) and within {+-}2.9%({sigma}{sub MC}=0.8%) in low-density lung ({rho}=0.1 g cm{sup -3}). In comparison, differences of up to 10.2% and 17.5% in lung and low-density lung were observed in the equivalent AAA calculations. Acuros XB dose calculations performed on a phantom containing an air cavity ({rho}=0.001 g cm{sup -3}) were found to be within the range of {+-}1.5% to {+-}4.5% of the BEAMnrc/DOSXYZnrc calculated benchmark ({sigma}{sub MC}=0.8%) in the tissue above and below the air cavity. A comparison of Acuros XB dose calculations performed on a lung CT dataset with a BEAMnrc/DOSXYZnrc benchmark shows agreement within {+-}2%/2mm and indicates that the remaining differences are primarily a result of differences in physical material assignments within a CT dataset. Conclusions: By considering the fundamental particle interactions in matter based on theoretical interaction cross sections, the Acuros XB algorithm is capable of modeling radiotherapy dose deposition with accuracy only previously achievable with Monte Carlo techniques.
- OSTI ID:
- 22096981
- Journal Information:
- Medical Physics, Vol. 38, Issue 4; Other Information: (c) 2011 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-E-T-290: Dosimetric Accuracy of Acuros XB and Analytical Anisotropic Algorithm in Stereotactic Ablative Radiotherapy of Small Lung Lesions
Accuracy of Acuros XB and AAA dose calculation for small fields with reference to RapidArc stereotactic treatments
Related Subjects
60 APPLIED LIFE SCIENCES
ACCURACY
AIR
ALGORITHMS
ALLOCATIONS
ANISOTROPY
BENCHMARKS
COMPARATIVE EVALUATIONS
COMPUTERIZED TOMOGRAPHY
CROSS SECTIONS
DATASETS
DOSIMETRY
FIELD TESTS
INTERACTIONS
INTERFACES
IONIZATION CHAMBERS
LUNGS
MONTE CARLO METHOD
PHANTOMS
PHOTONS
RADIATION DOSES
RADIOTHERAPY
SIMULATION
SKELETON
STANDARDS
VALIDATION