Title: Real-time verification of multileaf collimator-driven radiotherapy using a novel optical attenuation-based fluence monitor

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3549766· OSTI ID:22096933
; ;  [1]
  1. Departement de Radio-Oncologie et Centre de Recherche en Cancerologie, Ho circumflex tel-Dieu de Quebec, 11 Cote du Palais, Quebec, Quebec G1R 2J6 (Canada) and Departement de Physique, de Genie Physique et d'Optique, Universite Laval, Quebec, Quebec G1V 0A6 (Canada)

Purpose: Multileaf collimator (MLC)-driven conformal radiotherapy modalities [e.g., such as intensity-modulated radiotherapy (IMRT), intensity-modulated arc therapy, and stereotactic body radiotherapy] are more subject to delivery errors and dose calculation inaccuracies than standard modalities. Fluence monitoring during treatment delivery could reduce such errors by allowing an independent interface to quantify and assess measured difference between the delivered and planned treatment administration. We developed an optical attenuation-based detector to monitor fluence for the on-line quality control of radiotherapy delivery. The purpose of the current study was to develop the theoretical background of the invention and to evaluate the detector's performance both statistically and in clinical situations. Methods: We aligned 60 27-cm scintillating fibers coupled to a photodetector via clear optical fibers in the direction of motion of each of the 60 leaf pairs of a 120 leaves Millenium MLC on a Varian Clinac iX. We developed a theoretical model to predict the intensity of light collected on each side of the scintillating fibers when placed under radiation fields of varying sizes, intensities, and positions. The model showed that both the central position of the radiation field on the fiber (x{sub c}) and the integral fluence passing through the fiber ({Phi}{sub int}) could be assessed independently in a single measurement. We evaluated the performance of the prototype by (1) measuring the intrinsic variation of the measured values of x{sub c} and {Phi}{sub int}, (2) measuring the impact on the measured values of x{sub c} and {Phi}{sub int} of random leaf positioning errors introduced into IMRT fields, and (3) comparing the predicted values of x{sub c} and {Phi}{sub int} calculated with the treatment planning software to the measured values of x{sub c} and {Phi}{sub int} in order to assess the predictive effectiveness of the developed theoretical model. Results: We observed a very low intrinsic dispersion, dominated by Poisson statistics, for both x{sub c} (standard deviations of less than 1 mm) and {Phi}{sub int} (standard deviations of less than 0.20%). When confronted with random leaf positioning errors from IMRT segments, {Phi}{sub int} was highly sensitive to single leaf positioning errors as small as 1 mm at isocenter, while x{sub c} was sensitive to leaf pair translation errors of at least 2 mm at isocenter. Owing to the uncertainties in the doses calculated in regions of high perpendicular dose gradients, the measured values of x{sub c} and {Phi}{sub int} deviated from the predicted values of x{sub c} and {Phi}{sub int} by a mean of 1.3 mm and 2.6%, respectively. Conclusion: Our study showed that an optical attenuation-based detector can be used to effectively monitor integral fluence during radiotherapy delivery. The performance of such a system would enable real-time quality control of the incident fluence in current MLC-driven treatments such as IMRT and future adaptive radiotherapy procedures where new treatment plans will have to be delivered without passing thru the current standard quality control chain.

OSTI ID:
22096933
Journal Information:
Medical Physics, Vol. 38, Issue 3; Other Information: (c) 2011 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English