Direct measurement of absorbed dose to water in HDR {sup 192}Ir brachytherapy: Water calorimetry, ionization chamber, Gafchromic film, and TG-43
- Medical Physics Unit, McGill University, Montreal General Hospital, Montreal, Quebec H3G 1A4 (Canada)
Purpose: Gafchromic film and ionometric calibration procedures for HDR {sup 192}Ir brachytherapy sources in terms of dose rate to water are presented and the experimental results are compared to the TG-43 protocol as well as with the absolute dose measurement results from a water calorimetry-based primary standard. Methods: EBT-1 Gafchromic films, an A1SL Exradin miniature Shonka thimble type chamber, and an SI HDR 1000 Plus well-type chamber (Standard Imaging, Inc., Middleton, WI) with an ADCL traceable S{sub k} calibration coefficient (following the AAPM TG-43 protocol) were used. The Farmer chamber and Gafchromic film measurements were performed directly in water. All results were compared to direct and absolute absorbed dose to water measurements from a 4 deg. C stagnant water calorimeter. Results: Based on water calorimetry, the authors measured the dose rate to water to be 361{+-}7 {mu}Gy/(h U) at a 55 mm source-to-detector separation. The dose rate normalized to air-kerma strength for all the techniques agree with the water calorimetry results to within 0.83%. The overall 1-sigma uncertainty on water calorimetry, ionization chamber, Gafchromic film, and TG-43 dose rate measurement amounts to 1.90%, 1.44%, 1.78%, and 2.50%, respectively. Conclusions: This work allows us to build a more realistic uncertainty estimate for absorbed dose to water determination using the TG-43 protocol. Furthermore, it provides the framework necessary for a shift from indirect HDR {sup 192}Ir brachytherapy dosimetry to a more accurate, direct, and absolute measurement of absorbed dose to water.
- OSTI ID:
- 22096678
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 4 Vol. 37; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Measurement of absorbed dose-to-water for an HDR {sup 192}Ir source with ionization chambers in a sandwich setup
Development of a water calorimetry-based standard for absorbed dose to water in HDR {sup 192}Ir brachytherapy
Direct absorbed dose to water determination based on water calorimetry in scanning proton beam delivery
Journal Article
·
Sun Sep 15 00:00:00 EDT 2013
· Medical Physics
·
OSTI ID:22230769
Development of a water calorimetry-based standard for absorbed dose to water in HDR {sup 192}Ir brachytherapy
Journal Article
·
Thu Apr 15 00:00:00 EDT 2010
· Medical Physics
·
OSTI ID:22100579
Direct absorbed dose to water determination based on water calorimetry in scanning proton beam delivery
Journal Article
·
Thu Jul 15 00:00:00 EDT 2010
· Medical Physics
·
OSTI ID:22096713