Evaluation of delivered monitor unit accuracy of gated step-and-shoot IMRT using a two-dimensional detector array

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.3310806· OSTI ID:22096645
; ; ; ; ; ; ; ; ;  [1]
  1. Department of Radiation Oncology, Hallym University College of Medicine, Seoul, 431070 (Korea, Republic of) and Department of Biomedical Engineering, College of Medicine, Catholic University of Korea, Seoul 137701 (Korea, Republic of)

Purpose: To overcome the problem of organ motion in intensity-modulated radiation therapy (IMRT), gated IMRT is often used for the treatment of lung cancer. In this study, the authors investigated the accuracy of the delivered monitor units (MUs) from each segment during gated IMRT using a two-dimensional detector array for user-specific verification purpose. Methods: The authors planned a 6 MV photon, seven-port step-and-shoot lung IMRT delivery. The respiration signals for gated IMRT delivery were obtained from the one-dimensional moving phantom using the real-time position management (RPM) system (Varian Medical Systems, Palo Alto, CA). The beams were delivered using a Clinac iX (Varian Medical Systems, Palo Alto, CA) with the Millennium 120 MLC. The MatriXX (IBA Dosimetry GmbH, Germany) was validated through consistency and reproducibility tests as well as comparison with measurements from a Farmer-type ion chamber. The authors delivered beams with varying dose rates and duty cycles and analyzed the MatriXX data to evaluate MU delivery accuracy. Results: There was quite good agreement between the planned segment MUs and the MUs computed from the MatriXX within {+-}2% error. The beam-on times computed from the MatriXX data were almost identical for all cases, and they matched well with the RPM beam-on and beam-off signals. A slight difference was observed between them, but it was less than 40 ms. The gated IMRT delivery demonstrated an MU delivery accuracy that was equivalent to ungated IMRT, and the delivered MUs with a gating signal agreed with the planned MUs within {+-}0.5 MU regardless of dose rate and duty cycle. Conclusions: The authors can conclude that gated IMRT is able to deliver an accurate dose to a patient during a procedure. The authors believe that the methodology and results can be transferred to other vendors' devices, particularly those that do not provide MLC log data for a verification purpose.

OSTI ID:
22096645
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 3 Vol. 37; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Examination of geometric and dosimetric accuracies of gated step-and-shoot intensity modulated radiation therapy
Journal Article · Mon Oct 15 00:00:00 EDT 2007 · Medical Physics · OSTI ID:21032812

Monte Carlo based beam model using a photon MLC for modulated electron radiotherapy
Journal Article · Fri Feb 14 23:00:00 EST 2014 · Medical Physics · OSTI ID:22251146

Monte Carlo based beam model using a photon MLC for modulated electron radiotherapy
Journal Article · Fri Feb 14 23:00:00 EST 2014 · Medical Physics · OSTI ID:22251664