Development and performance evaluation of an experimental fine pitch detector multislice CT scanner
- MI and CT Engineering, GE Yokogawa Medical Systems, 4-7-127, Asahigaoka, Hino, Tokyo 191-8503 (Japan)
The authors have developed an experimental fine pitch detector multislice CT scanner with an ultrasmall focal spot x-ray tube and a high-density matrix detector through current CT technology. The latitudinal size of the x-ray tube focal spot was 0.4 mm. The detector dimension was 1824 channels (azimuthal direction)x32 rows (longitudinal direction) at row width of 0.3125 mm, in which a thinner reflected separator surrounds each detector cell coupled with a large active area photodiode. They were mounted on a commercial 64-slice CT scanner gantry while the scan field of view (50 cm) and gantry rotation speed (0.35 s) can be maintained. The experimental CT scanner demonstrated the spatial resolution of 0.21-0.22 mm (23.8-22.7 lp/cm) with the acrylic slit phantom and in-plane 50%-MTF 9.0 lp/cm and 10%-MTF 22.0 lp/cm. In the longitudinal direction, it demonstrated the spatial resolution of 0.24 mm with the high-resolution insert of the CATPHAN phantom and 0.34 mm as the full width at half maximum of the slice sensitivity profile. In low-contrast detectability, 3 mm at 0.3% was visualized at the CTDI{sub vol} of 47.2 mGy. Two types of 2.75 mm diameter vessel phantoms with in-stent stenosis at 25%, 50%, and 75% stair steps were scanned, and the reconstructed images can clearly resolve the stenosis at each case. The experimental CT scanner provides high-resolution imaging while maintaining low-contrast detectability, demonstrating the potentiality for clinical applications demanding high spatial resolution, such as imaging of inner ear, lung, and bone, or low-contrast detectability, such as imaging of coronary artery.
- OSTI ID:
- 22095313
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 4 Vol. 36; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Novel ultrahigh resolution data acquisition and image reconstruction for multi-detector row CT
Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot
SU-G-206-02: Impact of Focal Spot Sizes On CT Image Quality
Journal Article
·
Tue May 15 00:00:00 EDT 2007
· Medical Physics
·
OSTI ID:20951302
Image reconstruction and image quality evaluation for a 64-slice CT scanner with z-flying focal spot
Journal Article
·
Mon Aug 15 00:00:00 EDT 2005
· Medical Physics
·
OSTI ID:20726267
SU-G-206-02: Impact of Focal Spot Sizes On CT Image Quality
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22649307