Determination of absorbed dose in water at the reference point D(r{sub 0},{theta}{sub 0}) for an {sup 192}Ir HDR brachytherapy source using a Fricke system
- Department of Radiation Oncology, East Carolina University, Greenville, North Carolina 27834 (United States)
A ring-shaped Fricke device was developed to measure the absolute dose on the transverse bisector of a {sup 192}Ir high dose rate (HDR) source at 1 cm from its center in water, D(r{sub 0},{theta}{sub 0}). It consists of a polymethylmethacrylate (PMMA) rod (axial axis) with a cylindrical cavity at its center to insert the {sup 192}Ir radioactive source. A ring cavity around the source with 1.5 mm thickness and 5 mm height is centered at 1 cm from the central axis of the source. This ring cavity is etched in a disk shaped base with 2.65 cm diameter and 0.90 cm thickness. The cavity has a wall around it 0.25 cm thick. This ring is filled with Fricke solution, sealed, and the whole assembly is immersed in water during irradiations. The device takes advantage of the cylindrical geometry to measure D(r{sub 0},{theta}{sub 0}). Irradiations were performed with a Nucletron microselectron HDR unit loaded with an {sup 192}Ir Alpha Omega radioactive source. A Spectronic 1001 spectrophotometer was used to measure the optical absorbance using a 1 mL quartz cuvette with 1.00 cm light pathlength. The PENELOPE Monte Carlo code (MC) was utilized to simulate the Fricke device and the {sup 192}Ir Alpha Omega source in detail to calculate the perturbation introduced by the PMMA material. A NIST traceable calibrated well type ionization chamber was used to determine the air-kerma strength, and a published dose-rate constant was used to determine the dose rate at the reference point. The time to deliver 30.00 Gy to the reference point was calculated. This absorbed dose was then compared to the absorbed dose measured by the Fricke solution. Based on MC simulation, the PMMA of the Fricke device increases the D(r{sub 0},{theta}{sub 0}) by 2.0%. Applying the corresponding correction factor, the D(r{sub 0},{theta}{sub 0}) value assessed with the Fricke device agrees within 2.0% with the expected value with a total combined uncertainty of 3.43%(k=1). The Fricke device provides a promising method towards calibration of brachytherapy radiation sources in terms of D(r{sub 0},{theta}{sub 0}) and audit HDR source calibrations.
- OSTI ID:
- 22095275
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 12 Vol. 35; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
Quality assurance of HDR {sup 192}Ir sources using a Fricke dosimeter
SU-F-19A-02: Comparison of Absorbed Dose to Water Standards for HDR Ir-192 Brachytherapy Between the LCR, Brazil and NRC, Canada
SU-F-T-33: Air-Kerma Strength and Dose Rate Constant by the Full Monte Carlo Simulations
Journal Article
·
Sun Apr 15 00:00:00 EDT 2007
· Medical Physics
·
OSTI ID:20951152
SU-F-19A-02: Comparison of Absorbed Dose to Water Standards for HDR Ir-192 Brachytherapy Between the LCR, Brazil and NRC, Canada
Journal Article
·
Sun Jun 15 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22402286
SU-F-T-33: Air-Kerma Strength and Dose Rate Constant by the Full Monte Carlo Simulations
Journal Article
·
Wed Jun 15 00:00:00 EDT 2016
· Medical Physics
·
OSTI ID:22642283