The use of EPID-measured leaf sequence files for IMRT dose reconstruction in adaptive radiation therapy

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.2990782· OSTI ID:22095252
; ;  [1]
  1. Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California 94305 (United States)

For intensity modulated radiation treatment (IMRT) dose reconstruction, multileaf collimator (MLC) log files have been shown applicable for deriving delivered fluence maps. However, MLC log files are dependent on the accuracy of leaf calibration and only available from one linear accelerator manufacturer. This paper presents a proof of feasibility and principles in (1) using an amorphous silicon electronic portal imaging device (aSi-EPID) to capture the MLC segments during an IMRT delivery and (2) reconstituting a leaf sequence (LS) file based on the leaf end positions calculated from the MLC segments and their associated fractional monitor units. These EPID-measured LS files are then used to derive delivered fluence maps for dose reconstruction. The developed approach was tested on a pelvic phantom treated with a typical prostate IMRT plan. The delivered fluence maps, which were derived from the EPID-measured LS files, showed slight differences in the intensity levels compared with the corresponding planned ones. The dose distribution calculated with the delivered fluence maps showed a discernible difference in the high dose region when compared to that calculated with the planned fluence maps. The maximum dose in the former distribution was also 2.5% less than that in the latter one. The EPID-measured LS file can serve the same purpose as a MLC log files does for the derivation of the delivered fluence map and yet is independent of the leaf calibration. The approach also allows users who do not have access to MLC log files to probe the actual IMRT delivery and translate the information gained for dose reconstruction in adaptive radiation therapy.

OSTI ID:
22095252
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 11 Vol. 35; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

A high-speed scintillation-based electronic portal imaging device to quantitatively characterize IMRT delivery
Journal Article · Sat Jan 14 23:00:00 EST 2006 · Medical Physics · OSTI ID:20774962

SU-F-T-295: MLCs Performance and Patient-Specific IMRT QA Using Log File Analysis
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22648904

Retrospective IMRT Dose Reconstruction Based on Cone-Beam CT and MLC Log-File
Journal Article · Thu Jan 31 23:00:00 EST 2008 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21039776