Advances in compact proton spectrometers for inertial-confinement fusion and plasma nuclear science
Journal Article
·
· Review of Scientific Instruments
- Plasma Science and Fusion Center, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139 (United States)
- Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623 (United States)
Compact wedge-range-filter proton spectrometers cover proton energies {approx}3-20 MeV. They have been used at the OMEGA laser facility for more than a decade for measuring spectra of primary D{sup 3}He protons in D{sup 3}He implosions, secondary D{sup 3}He protons in DD implosions, and ablator protons in DT implosions; they are now being used also at the National Ignition Facility. The spectra are used to determine proton yields, shell areal density at shock-bang time and compression-bang time, fuel areal density, and implosion symmetry. There have been changes in fabrication and in analysis algorithms, resulting in a wider energy range, better accuracy and precision, and better robustness for survivability with indirect-drive inertial-confinement-fusion experiments.
- OSTI ID:
- 22093875
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 83; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
A novel particle Time Of Flight (pTOF) diagnostic for measurements of shock- and compression-bang times in D3He and DT implosions at the NIF
A novel particle time of flight diagnostic for measurements of shock- and compression-bang times in D{sup 3}He and DT implosions at the NIF
A magnetic particle time-of-flight (MagPTOF) diagnostic for measurements of shock- and compression-bang time at the NIF (invited)
Conference
·
Wed May 02 00:00:00 EDT 2012
·
OSTI ID:1043638
A novel particle time of flight diagnostic for measurements of shock- and compression-bang times in D{sup 3}He and DT implosions at the NIF
Journal Article
·
Mon Oct 15 00:00:00 EDT 2012
· Review of Scientific Instruments
·
OSTI ID:22093869
A magnetic particle time-of-flight (MagPTOF) diagnostic for measurements of shock- and compression-bang time at the NIF (invited)
Journal Article
·
Sun Jul 13 20:00:00 EDT 2014
· Review of Scientific Instruments
·
OSTI ID:1136977