Single-shot X-ray phase-contrast imaging using two-dimensional gratings
Journal Article
·
· AIP Conference Proceedings
- Frontier Research Center, Corporate R and D Headquarters, Canon Inc., 3-30-2 Shimomaruko, Ohta-ku, Tokyo 146-8501 (Japan)
We developed a two-dimensional gratings-based X-ray interferometer that requires only a single exposure for clinical radiography. The interferometer consisted of a checkerboard phase grating for {pi} phase modulation and a latticed amplitude grating. Using a synchrotron radiation source, the phase grating modulates the X-rays and generates a self-image, transformed to a moire fringe by the amplitude grating. To allow use of a conventional X-ray tube, the latticed source grating was installed downstream from the X-ray tube. Differential phase-contrast and scattering images in two orthogonal directions were obtained by Fourier analysis of the single moire fringe image and an absorption image. Results show that characteristic features of soft tissue in two orthogonal directions were clearly shown in the differential phase-contrast images.
- OSTI ID:
- 22069175
- Journal Information:
- AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 1466; ISSN APCPCS; ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
A grating-based single-shot x-ray phase contrast and diffraction method for in vivo imaging
Results from the first preclinical CT scanner with grating based phase contrast and a rotating gantry
Structured scintillator for hard x-ray grating interferometry
Journal Article
·
Sun Nov 14 23:00:00 EST 2010
· Medical Physics
·
OSTI ID:22096825
Results from the first preclinical CT scanner with grating based phase contrast and a rotating gantry
Journal Article
·
Tue Jul 31 00:00:00 EDT 2012
· AIP Conference Proceedings
·
OSTI ID:22069160
Structured scintillator for hard x-ray grating interferometry
Journal Article
·
Mon Apr 25 00:00:00 EDT 2011
· Applied Physics Letters
·
OSTI ID:21518402