Title: Obesity: An Independent Risk Factor for Insufficient Hemostasis Using the AngioSeal Vascular Closure Device After Antegrade Puncture

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2];  [1]
  1. University Hospital Saarland, Department of Diagnostic and Interventional Radiology (Germany)
  2. University Hospital Saarland, Institute of Medical Biometry, Epidemiology and Medical Informatics (Germany)

Purpose: This study was designed to investigate the efficacy of the AngioSeal vascular closure device after antegrade puncture of the femoral artery. Methods: In a prospective study, 120 consecutive patients underwent lower limb vascular intervention by an antegrade access to the common femoral artery (CFA). After intervention, a 6F (n = 88) or an 8F (n = 32) AngioSeal vascular closure device was used to achieve hemostasis. The technical success or the cause of failure was documented. In addition, the coagulation status (platelets, INR, prothrombin time, atrial thromboplastin time (PTT)), hypertonus, locoregional habitus of the groin, body mass index (BMI), presence of calcifications, and history of previous surgical interventions of the CFA were evaluated. Results: Hemostasis was achieved in 97 patients (81%). In 12 patients (10%), persistent bleeding of the puncture site required manual compression. In another nine patients (8%) a kink of the sheath obviated the passage of the collagen plug toward the vessel, and in two patients the anchor dislodged out of the vessel, requiring manual compression. There were no significant differences between the groups of successful and unsuccessful sealing regarding the mean platelets (241 vs. 254 * 10{sup 9}/l; P = 0.86), INR (1.06 vs. 1.02; P = 0.52), prothrombin time (90% vs. 90%; P = 0.86), and PTT (30 vs. 31 s; P = 0.82). However, unsuccessful sealing was more likely in obese patients with an increased BMI (26.6 vs. 28.8 kg/m{sup 2}; P = 0.04). Conclusions: Obesity seems to be an independent risk factor for insufficient sealing using the AngioSeal vascular closure device after antegrade puncture of the CFA. In 8% of our patients, hemostasis could not be achieved due to kink of the flexible sheath.

OSTI ID:
22066595
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 35, Issue 4; Other Information: Copyright (c) 2012 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Efficacy and Safety of the Angioseal Vascular Closure Device Post Antegrade Puncture
Journal Article · Thu May 15 00:00:00 EDT 2008 · Cardiovascular and Interventional Radiology · OSTI ID:22066595

Feasibility and Safety of Vascular Closure Devices in an Antegrade Approach to Either the Common Femoral Artery or the Superficial Femoral Artery
Journal Article · Mon Oct 15 00:00:00 EDT 2012 · Cardiovascular and Interventional Radiology · OSTI ID:22066595

Comparison of Ultrasound-Guided and Fluoroscopy-Assisted Antegrade Common Femoral Artery Puncture Techniques
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Cardiovascular and Interventional Radiology · OSTI ID:22066595