Title: Survival After Accidental Extrahepatic Distribution of Y90 Microspheres to the Mesentery During a Radioembolization Procedure

Journal Article · · Cardiovascular and Interventional Radiology
;  [1];  [2]; ;  [3];  [1]
  1. University Hospital, Department of Nuclear Medicine (Germany)
  2. University Hospital, Department of Surgery (Germany)
  3. University Hospital, Department of Radiology, Division of Radiation Oncology (Germany)

We present the acute management and outcome of a patient after an accidental mesenteric distribution of Y90 microspheres during radioembolization (RE). This report describes and highlights: (1) the incidence of a significant reflux during a RE session while injecting into a replaced right hepatic artery from the superior mesenteric artery, (2) the appearance of diffuse mesenteric Y90 distribution in bremsstrahlung-imaging, (3) the management protocol with the radiation protection agent amifostine, (4) the development of typical adverse effects in the expected time window, and (5) survival of the patient without long-term sequelae. This report should sensitize physicians to this particular problem and may help to avoid as well as manage similar radioembolization incidences.

OSTI ID:
22066581
Journal Information:
Cardiovascular and Interventional Radiology, Vol. 35, Issue 4; Other Information: Copyright (c) 2012 Springer Science+Business Media, LLC and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE); Country of input: International Atomic Energy Agency (IAEA); ISSN 0174-1551
Country of Publication:
United States
Language:
English

Similar Records

Angiographic Anatomy and Relevance of 3 and 9 O’clock Arteries During Radioembolization
Journal Article · Fri Jun 15 00:00:00 EDT 2018 · Cardiovascular and Interventional Radiology · OSTI ID:22066581

Safety and Efficacy of Y-90 Radioembolization After Prior Major Hepatic Resection
Journal Article · Tue Aug 15 00:00:00 EDT 2017 · Cardiovascular and Interventional Radiology · OSTI ID:22066581

Technical Solutions to Ensure Safe Yttrium-90 Radioembolization in Patients With Initial Extrahepatic Deposition of {sup 99m}Technetium-Albumin Macroaggregates
Journal Article · Sat Oct 15 00:00:00 EDT 2011 · Cardiovascular and Interventional Radiology · OSTI ID:22066581