Note: Multi-point measurement of |B| in the gas-dynamic trap with a spectral motional Stark effect diagnostic
Journal Article
·
· Review of Scientific Instruments
- Budker Institute of Nuclear Physics, 630090 Novosibirsk (Russian Federation)
- Department of Physics and Center for Magnetic Self-Organization in Laboratory and Astrophysical Plasmas, University of Wisconsin-Madison, Madison, Wisconsin 53706 (United States)
An upgraded spectral motional Stark effect diagnostic has been installed on the gas-dynamic trap (GDT) experiment to enable spatially resolved measurement of |B|. A new low-noise charge-coupled device detector, combined with enhancements of the diagnostic neutral beam, allows single-shot profile measurements. Previously only single-point motional Stark effect measurements were possible, and detector noise severely limited measurement precision, requiring multi-shot averaging. The plasma pressure profile in GDT is derived from the measured diamagnetic modification of |B| and used to examine the conditions of stable plasma confinement at high plasma pressure.
- OSTI ID:
- 22063708
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 8 Vol. 82; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Note: Spectral motional Stark effect diagnostic for measurement of magnetic fields below 0.3 T
Motional Stark Effect Diagnostic for Multi-Chord Measurements of Plasma Beta in GDT
Measurements of the internal magnetic field using the B-Stark motional Stark effect diagnostic on DIII-D (inivited)
Journal Article
·
Thu Aug 15 00:00:00 EDT 2013
· Review of Scientific Instruments
·
OSTI ID:22220478
Motional Stark Effect Diagnostic for Multi-Chord Measurements of Plasma Beta in GDT
Journal Article
·
Fri Jan 14 23:00:00 EST 2005
· Fusion Science and Technology
·
OSTI ID:20849808
Measurements of the internal magnetic field using the B-Stark motional Stark effect diagnostic on DIII-D (inivited)
Journal Article
·
Fri Oct 15 00:00:00 EDT 2010
· Review of Scientific Instruments
·
OSTI ID:22058691