Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Compact, rugged in-chamber transmission spectrometers (7-28 keV) for the Sandia Z facility

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3600610· OSTI ID:22062346
; ; ; ; ; ; ; ; ;  [1];  [2]
  1. Sandia National Laboratories, P.O. Box 5800, Albuquerque, New Mexico 87185 (United States)
  2. Laboratory of Plasma Studies, Cornell University, Ithaca, New York 14850 (United States)
We describe a pair of time-integrated transmission spectrometers that are designed to survey 7-28 keV (1.9 to 0.43 A) x-ray photons produced by experiments on the Sandia Z pulsed power facility. Each spectrometer uses a quartz 10-11 crystal in a Cauchois geometry with a slit to provide spatial resolution along one dimension. The spectrometers are located in the harsh environment of the Z vacuum chamber, which necessitates that their design be compact and rugged. Example data from calibration tests and Z experiments are shown that illustrate the utility of the instruments.
OSTI ID:
22062346
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 6 Vol. 82; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Compact rugged in-chamber transmission spectrometers (7-28 keV) for the Sandia Z facility.
Journal Article · Sat Oct 01 00:00:00 EDT 2011 · Review of Scientific Instruments · OSTI ID:1106020

Focusing, in-chamber spectrometer triplet for high resolution measurements on the Sandia Z facility
Journal Article · Sun Oct 15 00:00:00 EDT 2006 · Review of Scientific Instruments · OSTI ID:20861342

TIME-RESOLVED 1-10 keV CRYSTAL SPECTROMETER FOR THE Z MACHINE AT SANDIA NATIONAL LABORATORIES
Conference · Sun Jun 01 00:00:00 EDT 2003 · OSTI ID:811260