Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Electronic measurement of microchannel plate pulse height distributions

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3478687· OSTI ID:22055866
; ; ;  [1]
  1. University of Michigan, 2455 Hayward Street, Ann Arbor, Michigan 48109 (United States)
Microchannel plates are a central component of the x-ray framing cameras used as analog imagers in many plasma experiment diagnostic systems. The microchannel plate serves as an amplifying element, increasing the electronic signal from incident radiation by factors of 10{sup 3}-10{sup 5}, with a broad pulse-height distribution. Seeking to optimize the photon-to-electron conversion efficiency and noise distribution of x-ray cameras, we will characterize the pulse-height distribution of the electron output from a single microchannel plate. Replacing the framing camera's phosphor-coated fiber optic screen with a charge-collection plate and coupling to a low-noise multichannel analyzer, we quantified the distribution in the total charge generated per photon event. The electronically measured pulse height distribution is used to estimate the signal-to-noise ratio of radiographic images from framing cameras.
OSTI ID:
22055866
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 81; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English