A compact electron beam ion source with integrated Wien filter providing mass and charge state separated beams of highly charged ions
Journal Article
·
· Review of Scientific Instruments
- DREEBIT GmbH, Zur Wetterwarte 50, D-01109 Dresden (Germany)
- School of Nuclear Science and Technology, Lanzhou University, Lanzhou 730000 (China)
- Institut fuer Angewandte Physik, Technische Universitaet Dresden, Mommsenstr. 13, D-01062 Dresden (Germany)
- Institut fuer Theoretische Physik, Technische Universitaet Dresden, Mommsenstr. 13, D-01062 Dresden (Germany)
A Wien filter was designed for and tested with a room temperature electron beam ion source (EBIS). Xenon charge state spectra up to the charge state Xe{sup 46+} were resolved as well as the isotopes of krypton using apertures of different sizes. The complete setup consisting of an EBIS and a Wien filter has a length of less than 1 m substituting a complete classical beamline setup. The Wien filter is equipped with removable permanent magnets. Hence total beam current measurements are possible via simple removal of the permanent magnets. In dependence on the needs of resolution a weak (0.2 T) or a strong (0.5 T) magnets setup can be used. In this paper the principle of operation and the design of the Wien filter meeting the requirements of an EBIS are briefly discussed. The first ion beam extraction and separation experiments with a Dresden EBIS are presented.
- OSTI ID:
- 22053577
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 6 Vol. 80; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Compact electron beam ion sources/traps: Review and prospects (invited)
Liquid metal alloy ion source based metal ion injection into a room-temperature electron beam ion source
Wien filter with integrated vacuum pump
Journal Article
·
Thu Feb 14 23:00:00 EST 2008
· Review of Scientific Instruments
·
OSTI ID:21103945
Liquid metal alloy ion source based metal ion injection into a room-temperature electron beam ion source
Journal Article
·
Tue Feb 14 23:00:00 EST 2012
· Review of Scientific Instruments
·
OSTI ID:22063862
Wien filter with integrated vacuum pump
Patent
·
Tue Jul 06 00:00:00 EDT 2021
·
OSTI ID:1840237