Title: Prospective Study of Cone-Beam Computed Tomography Image-Guided Radiotherapy for Prone Accelerated Partial Breast Irradiation

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1]; ; ; ;  [1]
  1. Department of Radiation Oncology, New York University Medical Center, New York, NY (United States)

Purpose: To report setup variations during prone accelerated partial breast irradiation (APBI). Methods: New York University (NYU) 07-582 is an institutional review board-approved protocol of cone-beam computed tomography (CBCT) to deliver image-guided ABPI in the prone position. Eligible are postmenopausal women with pT1 breast cancer excised with negative margins and no nodal involvement. A total dose of 30 Gy in five daily fractions of 6 Gy are delivered to the planning target volume (the tumor cavity with 1.5-cm margin) by image-guided radiotherapy. Patients are set up prone, on a dedicated mattress, used for both simulation and treatment. After positioning with skin marks and lasers, CBCTs are performed and the images are registered to the planning CT. The resulting shifts (setup corrections) are recorded in the three principal directions and applied. Portal images are taken for verification. If they differ from the planning digital reconstructed radiographs, the patient is reset, and a new CBCT is taken. Results: 70 consecutive patients have undergone a total of 343 CBCTs: 7 patients had four of five planned CBCTs performed. Seven CBCTs (2%) required to be repeated because of misalignment in the comparison between portal and digital reconstructed radiograph image after the first CBCT. The mean shifts and standard deviations in the anterior-posterior (AP), superior-inferior (SI), and medial-lateral (ML) directions were -0.19 (0.54), -0.02 (0.33), and -0.02 (0.43) cm, respectively. The average root mean squares of the daily shifts were 0.50 (0.28), 0.29 (0.17), and 0.38 (0.20). A conservative margin formula resulted in a recommended margin of 1.26, 0.73, 0.96 cm in the AP, SI, and ML directions. Conclusion: CBCTs confirmed that the NYU prone APBI setup and treatment technique are reproducible, with interfraction variation comparable to those reported for supine setup. The currently applied margin (1.5 cm) adequately compensates for the setup variation detected.

OSTI ID:
21587739
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Vol. 81, Issue 2; Other Information: DOI: 10.1016/j.ijrobp.2010.11.029; PII: S0360-3016(10)03637-0; Copyright (c) 2011 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

Interfraction and Intrafraction Setup Variability for Prone Breast Radiation Therapy
Journal Article · Thu Apr 15 00:00:00 EDT 2010 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21587739

Generic Planning Target Margin for Rectal Cancer Treatment Setup Variation
Journal Article · Sat Aug 01 00:00:00 EDT 2009 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21587739

Comparison of Planned Versus Actual Dose Delivered for External Beam Accelerated Partial Breast Irradiation Using Cone-Beam CT and Deformable Registration
Journal Article · Mon Aug 01 00:00:00 EDT 2011 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21587739