Title: Effect of Embolization Material in the Calculation of Dose Deposition in Arteriovenous Malformations

Journal Article · · AIP Conference Proceedings
DOI:https://doi.org/10.1063/1.3531609· OSTI ID:21506980
 [1]; ;  [2]
  1. Unidad de Radioneurocirugia, Instituto Nacional de Neurologia y Neurocirugia. Insurgentes Sur 3677, Col. La Fama, C.P. 14269, Tlalpan, Mexico, D.F. (Mexico)
  2. Laboratorio de Fisica Medica, Instituto Nacional de Neurologia y Neurocirugia. Insurgentes Sur 3677, Col. La Fama, C.P. 14269, Tlalpan, Mexico, D.F. (Mexico)

In this work it is studied the impact of the incorporation of high Z materials (embolization material) in the dose calculation for stereotactic radiosurgery treatment for arteriovenous malformations. A statistical analysis is done to establish the variables that may impact in the dose calculation. To perform the comparison pencil beam (PB) and Monte Carlo (MC) calculation algorithms were used. The comparison between both dose calculations shows that PB overestimates the dose deposited. The statistical analysis, for the quantity of patients of the study (20), shows that the variable that may impact in the dose calculation is the volume of the high Z material in the arteriovenous malformation. Further studies have to be done to establish the clinical impact with the radiosurgery result.

OSTI ID:
21506980
Journal Information:
AIP Conference Proceedings, Vol. 1310, Issue 1; Conference: 11. Mexican symposium on medical physics, Mexico City (Mexico), 19-22 Mar 2010; Other Information: DOI: 10.1063/1.3531609; (c) 2010 American Institute of Physics; ISSN 0094-243X
Country of Publication:
United States
Language:
English