Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Inter- and Intrafraction Patient Positioning Uncertainties for Intracranial Radiotherapy: A Study of Four Frameless, Thermoplastic Mask-Based Immobilization Strategies Using Daily Cone-Beam CT

Journal Article · · International Journal of Radiation Oncology, Biology and Physics
 [1];  [2];  [1];  [3]; ; ; ;  [1]
  1. Department of Radiation Oncology and Molecular Radiation Sciences, Johns Hopkins University, School of Medicine, Baltimore, MD (United States)
  2. University of Maryland, School of Medicine, Baltimore, MD (United States)
  3. Johns Hopkins University, School of Medicine, Baltimore, MD (United States)

Purpose: To determine whether frameless thermoplastic mask-based immobilization is adequate for image-guided cranial radiosurgery. Methods and Materials: Cone-beam CT localization data from patients with intracranial tumors were studied using daily pre- and posttreatment scans. The systems studied were (1) Type-S IMRT (head only) mask (Civco) with head cushion; (2) Uni-Frame mask (Civco) with head cushion, coupled with a BlueBag body immobilizer (Medical Intelligence); (3) Type-S head and shoulder mask with head and shoulder cushion (Civco); (4) same as previous, coupled with a mouthpiece. The comparative metrics were translational shift magnitude and average rotation angle; systematic inter-, random inter-, and random intrafraction positioning error was computed. For strategies 1-4, respectively, the analysis for interfraction variability included data from 20, 9, 81, and 11 patients, whereas that for intrafraction variability included a subset of 7, 9, 16, and 8 patients. The results were compared for statistical significance using an analysis of variance test. Results: Immobilization system 4 provided the best overall accuracy and stability. The mean interfraction translational shifts ({+-} SD) were 2.3 ({+-} 1.4), 2.2 ({+-} 1.1), 2.7 ({+-} 1.5), and 2.1 ({+-} 1.0) mm whereas intrafraction motion was 1.1 ({+-} 1.2), 1.1 ({+-} 1.1), 0.7 ({+-} 0.9), and 0.7 ({+-} 0.8) mm for devices 1-4, respectively. No significant correlation between intrafraction motion and treatment time was evident, although intrafraction motion was not purely random. Conclusions: We find that all frameless thermoplastic mask systems studied are viable solutions for image-guided intracranial radiosurgery. With daily pretreatment corrections, symmetric PTV margins of 1 mm would likely be adequate if ideal radiation planning and targeting systems were available.

OSTI ID:
21491708
Journal Information:
International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 1 Vol. 80; ISSN IOBPD3; ISSN 0360-3016
Country of Publication:
United States
Language:
English

Similar Records

Evaluation of Image-Guided Positioning for Frameless Intracranial Radiosurgery
Journal Article · Wed Jul 01 00:00:00 EDT 2009 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:21276882

Six dimensional analysis with daily stereoscopic x-ray imaging of intrafraction patient motion in head and neck treatments using five points fixation masks
Journal Article · Tue Feb 14 23:00:00 EST 2006 · Medical Physics · OSTI ID:20775075

Evaluation of a Thermoplastic Immobilization System for Breast and Chest Wall Radiation Therapy
Journal Article · Fri Apr 01 00:00:00 EDT 2011 · Medical Dosimetry · OSTI ID:21490063