  1. Department of Health Physics, Institute of Radiation Medicine, Chinese Academy of Medical Sciences and Peking Union Medical College, Tianjin (China)
Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have been obtained experimentally. Measurement results are compared with that of the Monte Carlo calculations reported, and no difference has been found between them.
OSTI ID:
21486866
Journal Information:
Medical Dosimetry, Journal Name: Medical Dosimetry Journal Issue: 4 Vol. 35; ISSN 0958-3947
Country of Publication:
United States
Language:
English