Breath-Hold Target Localization With Simultaneous Kilovoltage/Megavoltage Cone-Beam Computed Tomography and Fast Reconstruction
Journal Article
·
· International Journal of Radiation Oncology, Biology and Physics
- Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim (Germany)
Purpose: Hypofractionated high-dose radiotherapy for small lung tumors has typically been based on stereotaxy. Cone-beam computed tomography and breath-hold techniques have provided a noninvasive basis for precise cranial and extracranial patient positioning. The cone-beam computed tomography acquisition time of 60 s, however, is beyond the breath-hold capacity of patients, resulting in respiratory motion artifacts. By combining megavoltage (MV) and kilovoltage (kV) photon sources (mounted perpendicularly on the linear accelerator) and accelerating the gantry rotation to the allowed limit, the data acquisition time could be reduced to 15 s. Methods and Materials: An Elekta Synergy 6-MV linear accelerator, with iViewGT as the MV- and XVI as the kV-imaging device, was used with a Catphan phantom and an anthropomorphic thorax phantom. Both image sources performed continuous image acquisition, passing an angle interval of 90{sup o} within 15 s. For reconstruction, filtered back projection on a graphics processor unit was used. It reconstructed 100 projections acquired to a 512 x 512 x 512 volume within 6 s. Results: The resolution in the Catphan phantom (CTP528 high-resolution module) was 3 lines/cm. The spatial accuracy was within 2-3 mm. The diameters of different tumor shapes in the thorax phantom were determined within an accuracy of 1.6 mm. The signal-to-noise ratio was 68% less than that with a 180{sup o}-kV scan. The dose generated to acquire the MV frames accumulated to 82.5 mGy, and the kV contribution was <6 mGy. Conclusion: The present results have shown that fast breath-hold, on-line volume imaging with a linear accelerator using simultaneous kV-MV cone-beam computed tomography is promising and can potentially be used for image-guided radiotherapy for lung cancer patients in the near future.
- OSTI ID:
- 21438059
- Journal Information:
- International Journal of Radiation Oncology, Biology and Physics, Journal Name: International Journal of Radiation Oncology, Biology and Physics Journal Issue: 4 Vol. 78; ISSN IOBPD3; ISSN 0360-3016
- Country of Publication:
- United States
- Language:
- English
Similar Records
Assessment of residual error in liver position using kV cone-beam computed tomography for liver cancer high-precision radiation therapy
On the use of EPID-based implanted marker tracking for 4D radiotherapy
Integrating respiratory gating into a megavoltage cone-beam CT system
Journal Article
·
Sun Oct 01 00:00:00 EDT 2006
· International Journal of Radiation Oncology, Biology and Physics
·
OSTI ID:20850142
On the use of EPID-based implanted marker tracking for 4D radiotherapy
Journal Article
·
Tue Nov 30 23:00:00 EST 2004
· Medical Physics
·
OSTI ID:20634540
Integrating respiratory gating into a megavoltage cone-beam CT system
Journal Article
·
Sat Jul 15 00:00:00 EDT 2006
· Medical Physics
·
OSTI ID:20853209